Revision of a blade cut-out in PFN-A fixation: Blade exchange, cement augmentation and a cement plug as a successful salvage option.
Trauma Case Rep
; 27: 100303, 2020 Jun.
Article
en En
| MEDLINE
| ID: mdl-32322652
ABSTRACT
A potential and feared complication of proximal femur nails with cephalomedullary fixation is migration of the cephalomedullary screw or blade (cut-out or cut-through). In patients not suitable (e.g. low demand, comorbidities) for conversion to total hip arthroplasty blade exchange with cement augmentation may be an option. This article describes the first successful clinical use of a salvage procedure of a previously published technique, which allows the surgeon to avoid intraarticular cement leakage by using a standard cement plug to close the defect in the femoral head.