Advanced design, simulation, and dosimetry of a novel rectal applicator for contact brachytherapy with a conventional HDR 192Ir source.
Bellezzo, Murillo; Fonseca, Gabriel P; Voncken, Robert; Verrijssen, An-Sofie; Van Beveren, Celine; Roelofs, Erik; Yoriyaz, Hélio; Reniers, Brigitte; Van Limbergen, Evert J; Berbée, Maaike; Verhaegen, Frank.
Afiliación
  • Bellezzo M; Department of Radiation Oncology (MAASTRO), GROW School for Oncology and Developmental Biology, Maastricht University Medical Center, Maastricht, The Netherlands; Centro de Engenharia Nuclear, Instituto de Pesquisas Energéticas e Nucleares IPEN-CNEN/SP, São Paulo, Brazil.
  • Fonseca GP; Department of Radiation Oncology (MAASTRO), GROW School for Oncology and Developmental Biology, Maastricht University Medical Center, Maastricht, The Netherlands.
  • Voncken R; Department of Radiation Oncology (MAASTRO), GROW School for Oncology and Developmental Biology, Maastricht University Medical Center, Maastricht, The Netherlands.
  • Verrijssen AS; Department of Radiation Oncology (MAASTRO), GROW School for Oncology and Developmental Biology, Maastricht University Medical Center, Maastricht, The Netherlands.
  • Van Beveren C; Department of Radiation Oncology (MAASTRO), GROW School for Oncology and Developmental Biology, Maastricht University Medical Center, Maastricht, The Netherlands.
  • Roelofs E; Department of Radiation Oncology (MAASTRO), GROW School for Oncology and Developmental Biology, Maastricht University Medical Center, Maastricht, The Netherlands.
  • Yoriyaz H; Centro de Engenharia Nuclear, Instituto de Pesquisas Energéticas e Nucleares IPEN-CNEN/SP, São Paulo, Brazil.
  • Reniers B; Research group NuTeC, Centre for Environmental Sciences, Hasselt University, Diepenbeek, Belgium.
  • Van Limbergen EJ; Department of Radiation Oncology (MAASTRO), GROW School for Oncology and Developmental Biology, Maastricht University Medical Center, Maastricht, The Netherlands.
  • Berbée M; Department of Radiation Oncology (MAASTRO), GROW School for Oncology and Developmental Biology, Maastricht University Medical Center, Maastricht, The Netherlands.
  • Verhaegen F; Department of Radiation Oncology (MAASTRO), GROW School for Oncology and Developmental Biology, Maastricht University Medical Center, Maastricht, The Netherlands. Electronic address: frank.verhaegen@maastro.nl.
Brachytherapy ; 19(4): 544-553, 2020.
Article en En | MEDLINE | ID: mdl-32386884
ABSTRACT

PURPOSE:

Dose escalation yields higher complete response to rectal tumors, which may enable the omission of surgery. Dose escalation using 50 kVp contact x-ray brachytherapy (CXB) allow the treatment of a selective volume, resulting in low toxicity and organs-at-risk preservation. However, the use of CXB devices is limited because of its high cost and lack of treatment planning tools. Hence, the MAASTRO applicator (for HDR 192Ir sources) was developed and characterized by measurements and Monte Carlo simulations to be a cost-effective alternative to CXB devices. METHODS AND MATERIALS A cylindrical applicator with lateral shielding was designed to be used with a rectoscope using its tip as treatment surface. Both the applicator and the rectoscope have a slanted edge to potentially allow easier placement against tumors. The applicator design was achieved by Monte Carlo modeling and validated experimentally with film dosimetry, using the Papillon 50 (P50) device as reference.

RESULTS:

The applicator delivers CXB doses in less than 9 min using a 20375 U source for a treatment area of approximately 20 × 20 mm2 at 2 mm depth. Normalized at 2 mm, the dose falloff for depths of 0 mm, 5 mm, and 10 mm are 130%, 70%, and 43% for the P50 and 140%, 67%, and 38% for the MAASTRO applicator, respectively.

CONCLUSIONS:

The MAASTRO applicator was designed to use HDR 192Ir sources to deliver a dose distribution similar to those of CXB devices. The applicator may provide a cost-effective solution for endoluminal boosting with clinical treatment planning system integration.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Neoplasias del Recto / Braquiterapia Tipo de estudio: Etiology_studies / Health_economic_evaluation Límite: Humans Idioma: En Revista: Brachytherapy Asunto de la revista: RADIOTERAPIA Año: 2020 Tipo del documento: Article País de afiliación: Brasil

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Neoplasias del Recto / Braquiterapia Tipo de estudio: Etiology_studies / Health_economic_evaluation Límite: Humans Idioma: En Revista: Brachytherapy Asunto de la revista: RADIOTERAPIA Año: 2020 Tipo del documento: Article País de afiliación: Brasil