Resolution enhancement in an extended depth of field for volumetric two-photon microscopy.
Opt Lett ; 45(11): 3054-3057, 2020 Jun 01.
Article en En | MEDLINE | ID: mdl-32479457
ABSTRACT
The resolution enhancement over the extended depth of field (DOF) in the volumetric two-photon microscopy (TPM) is demonstrated by utilizing multiple orders of Bessel beams. Here the conventional method of switching laser modes (SLAM) in 2D is introduced to 3D, denoted as the volumetric SLAM (V-SLAM). The equivalent scanning beam in the TPM is a thin needle-like beam, which is generated from the subtraction between the needle-like 0th-order and the straw-like 1st-order Bessel beams. Compared with the 0th-order Bessel beam, the lateral resolution of the V-SLAM is increased by 28.6% and maintains over the axial depth of 56 µm. The V-SLAM performance is evaluated by employing fluorescent beads and a mouse brain slice. The V-SLAM approach provides a promising solution to improve the lateral resolutions for fast volumetric imaging on sparsely distributed samples.

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Idioma: En Revista: Opt Lett Año: 2020 Tipo del documento: Article

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Idioma: En Revista: Opt Lett Año: 2020 Tipo del documento: Article
...