Comparing near infrared spectroscopy and transcutaneous oxygen measurement in hard-to-heal wounds: a pilot study.
Serena, Thomas E; Yaakov, Raphael; Serena, Laura; Mayhugh, Timothy; Harrell, Khristina.
Afiliación
  • Serena TE; SerenaGroup Research Foundation, Cambridge, MA, US.
  • Yaakov R; SerenaGroup Research Foundation, Cambridge, MA, US.
  • Serena L; SerenaGroup Research Foundation, Cambridge, MA, US.
  • Mayhugh T; SerenaGroup Research Foundation, Cambridge, MA, US.
  • Harrell K; SerenaGroup Research Foundation, Cambridge, MA, US.
J Wound Care ; 29(Sup6): S4-S9, 2020 Jun 01.
Article en En | MEDLINE | ID: mdl-32530756
ABSTRACT

OBJECTIVE:

Oxygen plays an integral role in all phases of the wound healing process and tissue oxygenation is a key determinant of wound healing. A comprehensive evaluation of patients with hard-to-heal wounds must include measurement of oxygenation in and around the area of skin breakdown. The current gold standard, transcutaneous oxygen measurement (TCOM), has numerous drawbacks and as a result has fallen into disuse.

METHOD:

This study compared measurement of tissue oxygenation of near infrared spectroscopy (NIRS) with TCOM in patients with acute and hard-to-heal wounds. The Shapiro-Wilk test was used to evaluate the normality of the data. The level of agreement between NIRS and TCOM was determined using Bland-Altman analysis. The relationship between TCOM and NIRS was examined using Pearson correlation.

RESULTS:

A total of 24 observations were obtained from 10 patients using TCOM and NIRS. The weighted mean partial pressure of oxygen (pO2) in the study population was 39.54mmHg (8.96 standard deviation). Bland-Altman analysis showed that mean difference was positive (18.75), suggesting an overestimation of oxygen measurements using TCOM compared with NIRS. The oxygen levels measured by TCOM and NIRS showed a strong correlation (r=0.74).

CONCLUSION:

The wound and hyperbaric community would benefit from a simplified procedure for measuring tissue oxygenation. These findings suggest a strong trend toward correlation between NIRS and TCOM. A further study in a larger population is recommended. NIRS offers several advantages over TCOM. Clinicians have immediate point-of-care visualisation of tissue oxygenation using a handheld device. The procedure takes minutes to perform and is less operator-dependent than TCOM. Finally, NIRS allows measurement of oxygenation in the wound bed, while TCOM does not.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Cicatrización de Heridas / Heridas y Lesiones / Monitoreo de Gas Sanguíneo Transcutáneo / Espectroscopía Infrarroja Corta / Oxigenoterapia Hiperbárica / Monitoreo Fisiológico Límite: Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: J Wound Care Asunto de la revista: ENFERMAGEM Año: 2020 Tipo del documento: Article País de afiliación: Estados Unidos

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Cicatrización de Heridas / Heridas y Lesiones / Monitoreo de Gas Sanguíneo Transcutáneo / Espectroscopía Infrarroja Corta / Oxigenoterapia Hiperbárica / Monitoreo Fisiológico Límite: Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: J Wound Care Asunto de la revista: ENFERMAGEM Año: 2020 Tipo del documento: Article País de afiliación: Estados Unidos