Your browser doesn't support javascript.
loading
Quantification of plasma remdesivir and its metabolite GS-441524 using liquid chromatography coupled to tandem mass spectrometry. Application to a Covid-19 treated patient.
Alvarez, Jean-Claude; Moine, Pierre; Etting, Isabelle; Annane, Djillali; Larabi, Islam Amine.
Afiliación
  • Alvarez JC; Laboratoire de Pharmacologie-Toxicologie, Université de Versailles Saint-Quentin en Yvelines, Inserm U-1173, Hôpital Raymond Poincaré, AP-HP, 104, Boulevard R. Poincaré, 92380, Garches, France.
  • Moine P; Intensive Care Unit, Paris-Saclay University (Versailles Saint-Quentin-en-Yvelines), Inserm U-1173, Raymond Poincaré Hospital, AP-HP, Garches, France.
  • Etting I; Department of Pharmacology and Toxicology, Paris-Saclay University (Versailles Saint-Quentin-en-Yvelines), Inserm U-1173, Raymond Poincaré Hospital, AP-HP, Garches, France.
  • Annane D; Intensive Care Unit, Paris-Saclay University (Versailles Saint-Quentin-en-Yvelines), Inserm U-1173, Raymond Poincaré Hospital, AP-HP, Garches, France.
  • Larabi IA; Department of Pharmacology and Toxicology, Paris-Saclay University (Versailles Saint-Quentin-en-Yvelines), Inserm U-1173, Raymond Poincaré Hospital, AP-HP, Garches, France.
Clin Chem Lab Med ; 58(9): 1461-1468, 2020 08 27.
Article en En | MEDLINE | ID: mdl-32573468
ABSTRACT

Objectives:

A method based on liquid chromatography coupled to triple quadrupole mass spectrometry detection using 50 µL of plasma was developed and fully validated for quantification of remdesivir and its active metabolites GS-441524.

Methods:

A simple protein precipitation was carried out using 75 µL of methanol containing the internal standard (IS) remdesivir-13C6 and 5 µL ZnSO4 1 M. After separation on Kinetex® 2.6 µm Polar C18 100A LC column (100 × 2.1 mm i.d.), both compounds were detected by a mass spectrometer with electrospray ionization in positive mode. The ion transitions used were m/z 603.3 â†’ m/z 200.0 and m/z 229.0 for remdesivir, m/z 292.2 â†’ m/z 173.1 and m/z 147.1 for GS-441524 and m/z 609.3 â†’ m/z 206.0 for remdesivir-13C6.

Results:

Calibration curves were linear in the 1-5000 µg/L range for remdesivir and 5-2500 for GS-441524, with limit of detection set at 0.5 and 2 µg/L and limit of quantification at 1 and 5 µg/L, respectively. Precisions evaluated at 2.5, 400 and 4000 µg/L for remdesivir and 12.5, 125, 2000 µg/L for GS-441524 were lower than 14.7% and accuracy was in the [89.6-110.2%] range. A slight matrix effect was observed, compensated by IS. Higher stability of remdesivir and metabolite was observed on NaF-plasma. After 200 mg IV single administration, remdesivir concentration decrease rapidly with a half-life less than 1 h while GS-441524 appeared rapidly and decreased slowly until H24 with a half-life around 12 h.

Conclusions:

This method would be useful for therapeutic drug monitoring of these compounds in Covid-19 pandemic.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Antivirales / Neumonía Viral / Pirroles / Triazinas / Adenosina Monofosfato / Cromatografía Liquida / Monitoreo de Drogas / Infecciones por Coronavirus / Alanina / Espectrometría de Masas en Tándem Límite: Female / Humans / Male / Middle aged Idioma: En Revista: Clin Chem Lab Med Asunto de la revista: QUIMICA CLINICA / TECNICAS E PROCEDIMENTOS DE LABORATORIO Año: 2020 Tipo del documento: Article País de afiliación: Francia

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Antivirales / Neumonía Viral / Pirroles / Triazinas / Adenosina Monofosfato / Cromatografía Liquida / Monitoreo de Drogas / Infecciones por Coronavirus / Alanina / Espectrometría de Masas en Tándem Límite: Female / Humans / Male / Middle aged Idioma: En Revista: Clin Chem Lab Med Asunto de la revista: QUIMICA CLINICA / TECNICAS E PROCEDIMENTOS DE LABORATORIO Año: 2020 Tipo del documento: Article País de afiliación: Francia