Simplified Approach to Detect Dielectric Constant Using a Low-Cost Microfluidic Quarter Mode Substrate-Integrated Waveguide.
Sensors (Basel)
; 20(17)2020 Sep 02.
Article
en En
| MEDLINE
| ID: mdl-32887520
ABSTRACT
Liquid materials' characterization using commercial probes and radio frequency techniques is expensive and complex. This study proposes a compact and cost-effective radio frequency sensor system to measure the dielectric constant using a three-material calibration. The simplified approach measures reflection coefficient magnitudes for all four materials rather than the complex values in conventional permittivity detection systems. We employ a sensor module based on a circular substrate-integrated waveguide with measured unloaded quality factor = 910 to ensure measurement reliability. Miniaturized quarter-mode substrate-integrated waveguide resonators are integrated with four microfluidic channels containing three known materials and one unknown analyte. Step-wise measurement and linearity ensures maximum 4% error for the dielectric constant compared with results obtained using a high-performance commercial product.