Your browser doesn't support javascript.
loading
A New Device for Fine-Needle Aspiration Cytology Consisting of a Vibrating Linear Resonant Actuator.
Taguchi, Atsushi; Kojima, Tsuyoshi; Okanoue, Yusuke; Kagoshima, Hiroki; Hasebe, Koki; Yamamoto, Hirotaka; Hori, Ryusuke.
Afiliación
  • Taguchi A; Department of Otolaryngology, Tenri Hospital, Nara, Japan.
  • Kojima T; Department of Otolaryngology, Head and Neck Surgery, Japan Red Cross Osaka Hospital, Osaka, Japan.
  • Okanoue Y; Department of Otolaryngology, Tenri Hospital, Nara, Japan.
  • Kagoshima H; Department of Otolaryngology, Tenri Hospital, Nara, Japan.
  • Hasebe K; Department of Otolaryngology, Tenri Hospital, Nara, Japan.
  • Yamamoto H; Department of Otolaryngology, Head and Neck Surgery, Graduate School of Medicine, Kyoto University, Kyoto, Japan.
  • Hori R; Department of Otolaryngology, Tenri Hospital, Nara, Japan.
Laryngoscope ; 131(4): E1393-E1399, 2021 04.
Article en En | MEDLINE | ID: mdl-33001466
ABSTRACT

OBJECTIVE:

To evaluate the efficacy of a new device for fine-needle aspiration cytology (FNAC) consisting of a vibrating linear resonant actuator (LRA). STUDY

DESIGN:

Prospective clinical study.

METHODS:

The LRA frequency was optimized by visualization of the needle motion using a high-speed camera. The FNAC device consists of a vibrating motor fixed to the stopper of a 5-ml syringe and piston. Upon insertion of the syringe needle into a thyroid nodule (with the stopper attached to the syringe piston), sufficient negative pressure with 1-ml suction was maintained. Subsequently, samples were obtained using vibration generated by an LRA or an eccentric rotating mass (ERM). Surgically resected thyroid specimens from 10 patients were evaluated. The number of follicular groups required for adequate diagnosis and the number of larger follicular groups were counted. Next, 254 thyroid nodules from 187 patients were also evaluated by FNAC. The inadequacy rate was determined, and final cytology was classified according to thyroid Bethesda categories.

RESULTS:

The optimized LRA frequency was 155 Hz. Both the LRA and ERM devices resulted in sufficient amounts of diagnostic material and achieved low inadequacy rates. The number of large follicular groups obtained was significantly greater with the LRA device compared with the ERM device.

CONCLUSIONS:

The vibrating device using an LRA for thyroid FNAC resulted in sufficient amounts of thyroid follicular groups and achieved low inadequacy rates. In addition, the LRA device allowed for collection of larger follicular groups sufficient to diagnose appropriate thyroid Bethesda categories. LEVEL OF EVIDENCE 2 Laryngoscope, 131E1393-E1399, 2021.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Vibración / Neoplasias de la Tiroides / Nódulo Tiroideo / Biopsia con Aguja Fina Tipo de estudio: Diagnostic_studies / Observational_studies Límite: Aged / Female / Humans / Male / Middle aged Idioma: En Revista: Laryngoscope Asunto de la revista: OTORRINOLARINGOLOGIA Año: 2021 Tipo del documento: Article País de afiliación: Japón

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Vibración / Neoplasias de la Tiroides / Nódulo Tiroideo / Biopsia con Aguja Fina Tipo de estudio: Diagnostic_studies / Observational_studies Límite: Aged / Female / Humans / Male / Middle aged Idioma: En Revista: Laryngoscope Asunto de la revista: OTORRINOLARINGOLOGIA Año: 2021 Tipo del documento: Article País de afiliación: Japón
...