Your browser doesn't support javascript.
loading
Analytical validation of eight methods of thyroglobulin measurement in fine-needle aspiration washouts.
Boux de Casson, Florence; Beloeil, Rémi; Gauchez, Anne-Sophie; Oris, Charlotte; Leban, Monique; Schlageter, Marie-Hélène; Moineau, Marie-Pierre; Dufour-Rainfray, Diane; Bach-Ngohou, Kalyane; Chikh, Karim; Moal, Valérie.
Afiliación
  • Boux de Casson F; Service de Biochimie et Biologie Moléculaire, Centre Hospitalier Universitaire, Angers, France.
  • Beloeil R; Groupe de Biologie Spécialisée de la Société Française de Médecine Nucléaire, Paris, France.
  • Gauchez AS; Laboratoire IHE/BMGS et Service DEL/DIS, Etablissement Français du Sang Bretagne, Brest, France.
  • Oris C; Groupe de Biologie Spécialisée de la Société Française de Médecine Nucléaire, Paris, France.
  • Leban M; Plateforme de Radioactivité, Centre Hospitalier Universitaire Grenoble - Alpes, UMR-S INSERM 1039, Grenoble, France.
  • Schlageter MH; Service de Médecine Nucléaire, Centre Hospitalier Métropole Savoie, Chambéry, France.
  • Moineau MP; Service de Biochimie, Centre Hospitalier Universitaire, Clermont-Ferrand, France.
  • Dufour-Rainfray D; Groupe de Biologie Spécialisée de la Société Française de Médecine Nucléaire, Paris, France.
  • Bach-Ngohou K; Département de Biochimie Hormonale, Centre Hospitalier Universitaire Pitié-Salpêtrière, AP-HP, Paris, France.
  • Chikh K; Groupe de Biologie Spécialisée de la Société Française de Médecine Nucléaire, Paris, France.
  • Moal V; Laboratoire de Biologie Cellulaire, Hôpital Saint-Louis, AP-HP, Paris, France.
Ann Clin Biochem ; 58(1): 54-65, 2021 01.
Article en En | MEDLINE | ID: mdl-33026828
ABSTRACT

BACKGROUND:

Thyroglobulin (Tg) assay in washout fluids of fine needles, after cervical lymph nodes aspiration, is used for detecting metastases from differentiated thyroid carcinomas. Assay methods are the same as for Tg in serum. However, with non-serum samples, methods require extensive validation to notably check for the absence of matrix effect. This study fits this context. Our objectives were to assess analytic performances, in washout fluid, of eight different Tg assay methods and to compare them to validated data in serum.

METHODS:

Eleven medical laboratories participated in this study. The matrix tested was phosphate-buffer saline containing 1% bovine serum albumin (PBS-1% BSA). Samples used were dilutions, in this buffer, of Certified Reference Material (CRM 457). We verified, for all methods, the limit of detection, precision, linearity, trueness and accuracy.

RESULTS:

In PBS-1% BSA, the functional sensitivities (FS) were comparable to those expected for serum. All the methods were linear. The relative biases of trueness were between -24.5 and 10.2% around 1 µg/L. Total analytical error was ≤40% near the functional sensitivity values.

CONCLUSION:

No quantitatively important matrix effect was observed. All the methods showed their ability to measure Tg in PBS-1% BSA, over the concentration range of interest, with acceptable total analytical error. We validated the functional sensitivity value as a decision threshold in thyroidectomized patients after treatment and with low concentrations of serum Tg.