Combined analysis of plasma or serum glucosylsphingosine and globotriaosylsphingosine by UPLC-MS/MS.
Beasley, James; McCaw, Patricia; Zhang, Haoyue; Young, Sarah P; Stiles, Ashlee R.
Afiliación
  • Beasley J; Biochemical Genetics Laboratory, Duke University Health System, Durham, NC, USA.
  • McCaw P; Biochemical Genetics Laboratory, Duke University Health System, Durham, NC, USA.
  • Zhang H; Biochemical Genetics Laboratory, Duke University Health System, Durham, NC, USA.
  • Young SP; Biochemical Genetics Laboratory, Duke University Health System, Durham, NC, USA; Department of Pediatrics, Duke University Medical Center, Durham, NC, USA.
  • Stiles AR; Biochemical Genetics Laboratory, Duke University Health System, Durham, NC, USA; Department of Pediatrics, Duke University Medical Center, Durham, NC, USA. Electronic address: Ashlee.Stiles@duke.edu.
Clin Chim Acta ; 511: 132-137, 2020 Dec.
Article en En | MEDLINE | ID: mdl-33058838
ABSTRACT

PURPOSE:

To develop a method for the combined analysis of plasma and serum glucosylsphingosine (lyso-Gb1) and globotriaosylsphingosine (lyso-Gb3), biomarkers of Gaucher disease (GD) and Fabry disease (FD), respectively.

METHODS:

Internal standards were added to 100 µL of plasma/serum and glycosphingolipids lyso-Gb1, lyso-Gb3, and galactosylsphingosine (GalSph) were extracted with dichloromethane/methanol and analyzed by UPLC-MS/MS. Samples from unaffected controls and patients with GD were first analyzed using a HILIC column to separate lyso-Gb1 from its isomer, GalSph. Samples from patients with FD or GD were analyzed using a C18 column to measure lyso-Gb3 and the hexosylsphingosine (HexSph lyso-Gb1 + GalSph) fraction in a single combined method.

RESULTS:

Extraction efficiency was between 73% and 87% and day-to-day variability showed a relative standard error of <7.5%. GalSph was determined to have minimal to no contribution to the HexSph fraction in samples from unaffected controls and patients with GD. Lyso-Gb3 and HexSph measurements by the combined method were in good agreement with established methods, with no bias.

CONCLUSIONS:

HexSph and lyso-Gb3 analysis by reversed-phase chromatography UPLC-MS/MS is a cost-effective, time-efficient approach for evaluating these glycosphingolipid biomarkers in patients with a suspected or confirmed diagnosis of GD and FD.