Real-time motion and retrospective coil sensitivity correction for CEST using volumetric navigators (vNavs) at 7T.
Poblador Rodriguez, Esau; Moser, Philipp; Auno, Sami; Eckstein, Korbinian; Dymerska, Barbara; van der Kouwe, Andre; Gruber, Stephan; Trattnig, Siegfried; Bogner, Wolfgang.
Afiliación
  • Poblador Rodriguez E; High Field MR Center, Department of Biomedical Imaging and Image-guided Therapy, Medical University Vienna, Vienna, Austria.
  • Moser P; High Field MR Center, Department of Biomedical Imaging and Image-guided Therapy, Medical University Vienna, Vienna, Austria.
  • Auno S; Neuroscience Center, University of Helsinki, Helsinki, Finland.
  • Eckstein K; High Field MR Center, Department of Biomedical Imaging and Image-guided Therapy, Medical University Vienna, Vienna, Austria.
  • Dymerska B; Medical Physics and Bioengineering, University College London, London, United Kingdom.
  • van der Kouwe A; Athinoula A. Martinos Center for Biomedical Imaging, Department of Radiology, Massachusetts General Hospital, Harvard Medical School, Boston, Massachusetts, USA.
  • Gruber S; High Field MR Center, Department of Biomedical Imaging and Image-guided Therapy, Medical University Vienna, Vienna, Austria.
  • Trattnig S; High Field MR Center, Department of Biomedical Imaging and Image-guided Therapy, Medical University Vienna, Vienna, Austria.
  • Bogner W; Christian Doppler Laboratory for Clinical Molecular MR Imaging, Vienna, Austria.
Magn Reson Med ; 85(4): 1909-1923, 2021 04.
Article en En | MEDLINE | ID: mdl-33165952
ABSTRACT

PURPOSE:

To explore the impact of temporal motion-induced coil sensitivity changes on CEST-MRI at 7T and its correction using interleaved volumetric EPI navigators, which are applied for real-time motion correction.

METHODS:

Five healthy volunteers were scanned via CEST. A 4-fold correction pipeline allowed the mitigation of (1) motion, (2) motion-induced coil sensitivity variations, ΔB1- , (3) motion-induced static magnetic field inhomogeneities, ΔB0 , and (4) spatially varying transmit RF field fluctuations, ΔB1+ . Four CEST measurements were performed per session. For the first 2, motion correction was turned OFF and then ON in absence of voluntary motion, whereas in the other 2 controlled head rotations were performed. During post-processing ΔB1- was removed additionally for the motion-corrected cases, resulting in a total of 6 scenarios to be compared. In all cases, retrospective ∆B0 and - ΔB1+ corrections were performed to compute artifact-free magnetization transfer ratio maps with asymmetric analysis (MTRasym ).

RESULTS:

Dynamic ΔB1- correction successfully mitigated signal deviations caused by head motion. In 2 frontal lobe regions of volunteer 4, induced relative signal errors of 10.9% and 3.9% were reduced to 1.1% and 1.0% after correction. In the right frontal lobe, the motion-corrected MTRasym contrast deviated 0.92%, 1.21%, and 2.97% relative to the static case for Δω = 1, 2, 3 ± 0.25 ppm. The additional application of ΔB1- correction reduced these deviations to 0.10%, 0.14%, and 0.42%. The fully corrected MTRasym values were highly consistent between measurements with and without intended head rotations.

CONCLUSION:

Temporal ΔB1- cause significant CEST quantification bias. The presented correction pipeline including the proposed retrospective ΔB1- correction significantly reduced motion-related artifacts on CEST-MRI.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Algoritmos / Procesamiento de Imagen Asistido por Computador Tipo de estudio: Diagnostic_studies / Observational_studies Límite: Humans Idioma: En Revista: Magn Reson Med Asunto de la revista: DIAGNOSTICO POR IMAGEM Año: 2021 Tipo del documento: Article País de afiliación: Austria

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Algoritmos / Procesamiento de Imagen Asistido por Computador Tipo de estudio: Diagnostic_studies / Observational_studies Límite: Humans Idioma: En Revista: Magn Reson Med Asunto de la revista: DIAGNOSTICO POR IMAGEM Año: 2021 Tipo del documento: Article País de afiliación: Austria
...