Comparison of the StaRRsed Interliner device with Westergren method in erythrocyte sedimentation rate measurement.
Orkmez, Mustafa; Orhan, Seren; Bozdayi, Mehmet Akif; Tarakcioglu, Mehmet.
Afiliación
  • Orkmez M; Department of Medical Biochemistry, Faculty of Medicine, Gaziantep University, Gaziantep, Turkey.
  • Orhan S; Department of Medical Biochemistry, Faculty of Medicine, Gaziantep University, Gaziantep, Turkey.
  • Bozdayi MA; Department of Medical Biochemistry, Faculty of Medicine, Gaziantep University, Gaziantep, Turkey.
  • Tarakcioglu M; Department of Medical Biochemistry, Faculty of Medicine, Gaziantep University, Gaziantep, Turkey.
Int J Lab Hematol ; 43(4): 616-622, 2021 Aug.
Article en En | MEDLINE | ID: mdl-33522096
ABSTRACT

INTRODUCTION:

With recent advances in technology, many manual tests are being replaced by automated devices due to a wide range of advantages. One of these tests is the erythrocyte sedimentation rate (ESR) test that is used to determine inflammatory activity. This study aimed to evaluate the agreement between the Starrsed Interliner sedimentation device and the gold standard method, that is the Westergren method, used in ESR measurement.

METHODS:

One hundred fifty-one patients who presented to Gaziantep University Faculty of Medicine, Sahinbey Training and Research Hospital were included in this study. ESR values were measured simultaneously within 2 hours using the ESR analyzer Starrsed Interliner device and the gold standard method of measuring ESR, that is the Westergren method, from blood samples collected from the same patients in EDTA and citrate tubes.

RESULTS:

Agreement between the results from the Starrsed Interliner device and the Westergren method was evaluated using the Intraclass Correlation method. Consequently, a poor correlation was observed at values <20 mm/h, a moderate correlation was observed at values 20 to 80 mm/h and >80 mm/h, and an excellent correlation was observed when all results were considered. Method comparison was conducted according to the Passing-Bablok regression analysis (y = -1.50 + 0.75x) (P < .0001). The mean difference between the two methods was 10.1 according to the Bland-Altman analysis.

CONCLUSIONS:

Despite the advantages of the Starrsed Interliner device, such as lower laboratory workloads, lower costs and turnaround time, the difference between the two methods, as found in this study, may lead to different clinical interpretations for results in some patient.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Sedimentación Sanguínea / Pruebas Diagnósticas de Rutina Tipo de estudio: Guideline Límite: Adolescent / Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: Int J Lab Hematol Asunto de la revista: HEMATOLOGIA Año: 2021 Tipo del documento: Article País de afiliación: Turquía Pais de publicación: ENGLAND / ESCOCIA / GB / GREAT BRITAIN / INGLATERRA / REINO UNIDO / SCOTLAND / UK / UNITED KINGDOM

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Sedimentación Sanguínea / Pruebas Diagnósticas de Rutina Tipo de estudio: Guideline Límite: Adolescent / Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: Int J Lab Hematol Asunto de la revista: HEMATOLOGIA Año: 2021 Tipo del documento: Article País de afiliación: Turquía Pais de publicación: ENGLAND / ESCOCIA / GB / GREAT BRITAIN / INGLATERRA / REINO UNIDO / SCOTLAND / UK / UNITED KINGDOM