The adaptation and investigation of cone-beam CT reconstruction algorithms for horizontal rotation fixed-gantry scans of rabbits.
Gardner, Mark; Dillon, Owen; Shieh, Chun-Chien; O'Brien, Ricky; Debrot, Emily; Barber, Jeffrey; Ahern, Verity; Bennett, Peter; Heng, Soo-Min; Corde, Stéphanie; Jackson, Michael; Keall, Paul.
Afiliación
  • Gardner M; ACRF Image X Institute, The University of Sydney, Eveleigh, NSW 2015, Australia.
  • Dillon O; ACRF Image X Institute, The University of Sydney, Eveleigh, NSW 2015, Australia.
  • Shieh CC; ACRF Image X Institute, The University of Sydney, Eveleigh, NSW 2015, Australia.
  • O'Brien R; Sydney Neuroimaging Analysis Centre, Camperdown, NSW 2050, Australia.
  • Debrot E; ACRF Image X Institute, The University of Sydney, Eveleigh, NSW 2015, Australia.
  • Barber J; ACRF Image X Institute, The University of Sydney, Eveleigh, NSW 2015, Australia.
  • Ahern V; Western Sydney Local Health District, Blacktown, NSW 2148, Australia.
  • Bennett P; Western Sydney Local Health District, Blacktown, NSW 2148, Australia.
  • Heng SM; Faculty of Science, The University of Sydney, Camperdown, NSW 2006, Australia.
  • Corde S; Nelune Comprehensive Cancer Centre, Randwick, NSW 2031, Australia.
  • Jackson M; Nelune Comprehensive Cancer Centre, Randwick, NSW 2031, Australia.
  • Keall P; Nelune Comprehensive Cancer Centre, Randwick, NSW 2031, Australia.
Phys Med Biol ; 66(10)2021 05 14.
Article en En | MEDLINE | ID: mdl-33878747
ABSTRACT
Fixed-gantry radiation therapy has been proposed as a low-cost alternative to the conventional rotating-gantry radiation therapy, that may help meet the rising global treatment demand. Fixed-gantry systems require gravitational motion compensated reconstruction algorithms to produce cone-beam CT (CBCT) images of sufficient quality for image guidance. The aim of this work was to adapt and investigate five CBCT reconstruction algorithms for fixed-gantry CBCT images. The five algorithms investigated were Feldkamp-Davis-Kress (FDK), prior image constrained compressed sensing (PICCS), gravitational motion compensated FDK (GMCFDK), motion compensated PICCS (MCPICCS) (a novel CBCT reconstruction algorithm) and simultaneous motion estimation and iterative reconstruction (SMEIR). Fixed-gantry and rotating-gantry CBCT scans were acquired of 3 rabbits, with the rotating-gantry scans used as a reference. Projections were sorted into rotation bins, based on the angle of rotation of the rabbit during image acquisition. The algorithms were compared using the structural similarity index measure root mean square error, and reconstruction time. Evaluation of the reconstructed volumes showed that, when compared with the reference rotating-gantry volume, the conventional FDK algorithm did not accurately reconstruct fixed-gantry CBCT scans. Whilst the PICCS reconstruction algorithm reduced some motion artefacts, the motion estimation reconstruction methods (GMCFDK, MCPICCS and SMEIR) were able to greatly reduce the effect of motion artefacts on the reconstructed volumes. This finding was verified quantitatively, with GMCFDK, MCPICCS and SMEIR reconstructions having RMSE 17%-19% lower and SSIM 1% higher than a conventional FDK. However, all motion compensated fixed-gantry CBCT reconstructions had a 56%-61% higher RMSE and 1.5% lower SSIM than FDK reconstructions of conventional rotating-gantry CBCT scans. The results show that motion compensation is required to reduce motion artefacts for fixed-gantry CBCT reconstructions. This paper further demonstrates the feasibility of fixed-gantry CBCT scans, and the ability of CBCT reconstruction algorithms to compensate for motion due to horizontal rotation.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Tomografía Computarizada de Haz Cónico / Tomografía Computarizada Cuatridimensional Tipo de estudio: Guideline Límite: Animals Idioma: En Revista: Phys Med Biol Año: 2021 Tipo del documento: Article País de afiliación: Australia

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Tomografía Computarizada de Haz Cónico / Tomografía Computarizada Cuatridimensional Tipo de estudio: Guideline Límite: Animals Idioma: En Revista: Phys Med Biol Año: 2021 Tipo del documento: Article País de afiliación: Australia