In vitro modeling accurately predicts cardiac lead fracture at 10 years.
Wilkoff, Bruce L; Donnellan, Eoin; Himes, Adam; Ben Johnson, W; Haddad, Tarek; Lulic, Thomas; Lexcen, Daniel R; Crossley, George H.
Afiliación
  • Wilkoff BL; Department of Cardiovascular Medicine, Cleveland Clinic, Cleveland, Ohio. Electronic address: wilkofb@ccf.org.
  • Donnellan E; Department of Cardiovascular Medicine, Cleveland Clinic, Cleveland, Ohio.
  • Himes A; Medtronic, plc, Mounds View, Minnesota.
  • Ben Johnson W; Iowa Heart Center PC, Des Moines, Iowa.
  • Haddad T; Medtronic, plc, Mounds View, Minnesota.
  • Lulic T; Medtronic, plc, Mounds View, Minnesota.
  • Lexcen DR; Medtronic, plc, Mounds View, Minnesota.
  • Crossley GH; Vanderbilt Heart and Vascular Institute, Nashville, Tennessee.
Heart Rhythm ; 18(9): 1605-1612, 2021 09.
Article en En | MEDLINE | ID: mdl-33992730
ABSTRACT

BACKGROUND:

Development of a cardiac lead fracture model has the potential to differentiate well-performing lead designs from poor performing ones and could aid in future lead development.

OBJECTIVE:

The purpose of this study was to demonstrate a predictive model for lead fracture and validate the results generated by the model by comparing them to observed 10-year implantable cardioverter-defibrillator lead fracture-free survival.

METHODS:

The model presented here uses a combination of in vivo patient data, in vitro conductor fatigue test data, and statistical simulation to predict the fracture-free survival of cardiac leads. The model was validated by comparing the results to human clinical performance data from the Medtronic Sprint Fidelis (Minneapolis, MN) models 6931 (single coil, active fixation) and 6949 (dual coil, active fixation), as well as the Quattro model 6947 (dual coil, active fixation).

RESULTS:

Median patient age in the single coil Fidelis 6931 population (64 years) was less than in the dual coil Fidelis 6949 and Quattro populations (68 years). Modeled and observed fracture-free survival for Quattro (>97%) was superior to that for Fidelis (<94%). The modeled survival agreed with the observed fracture-free survival data. The average model error was 0.3% (SD 1.2%).

CONCLUSION:

This model for cardiac lead fracture-free survival using in vivo lead bending measurements and in vitro bench testing can be used to predict lead performance as observed by alignment with field survival data.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Falla de Prótesis / Desfibriladores Implantables / Electrodos Implantados / Falla de Equipo Tipo de estudio: Etiology_studies / Prognostic_studies / Risk_factors_studies Límite: Female / Humans / Male / Middle aged Idioma: En Revista: Heart Rhythm Año: 2021 Tipo del documento: Article

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Falla de Prótesis / Desfibriladores Implantables / Electrodos Implantados / Falla de Equipo Tipo de estudio: Etiology_studies / Prognostic_studies / Risk_factors_studies Límite: Female / Humans / Male / Middle aged Idioma: En Revista: Heart Rhythm Año: 2021 Tipo del documento: Article
...