Guiding vacuum-assisted biopsy in prone position: digital breast tomosynthesis vs stereotactic.
Scaperrotta, Gianfranco P; Boffelli, Giulia; Depretto, Catherine; Di Leo, Giovanni; Liguori, Alessandro; Monaco, Cristian Giuseppe; Borelli, Anna; Ferranti, Claudio.
Afiliación
  • Scaperrotta GP; IRCCS Istituto Nazionale dei Tumori, Italy.
  • Boffelli G; ASST Papa Giovanni XXXIII, Bergamo, Italy.
  • Depretto C; IRCCS Istituto Nazionale dei Tumori, Italy.
  • Di Leo G; IRCCS Policlinico San Donato, Italy.
  • Liguori A; Policlinico di Milano, Italy.
  • Monaco CG; IRCCS Policlinico San Donato, Italy.
  • Borelli A; Università degli Studi di Milano, Italy.
  • Ferranti C; IRCCS Istituto Nazionale dei Tumori, Italy.
Tumori ; 108(4): 326-330, 2022 Aug.
Article en En | MEDLINE | ID: mdl-34041978
ABSTRACT

PURPOSE:

To compare the performance of prone digital breast tomosynthesis (DBT)-vacuum-assisted biopsy (VAB) with prone stereotactic-guided VAB (sVAB), focusing on time of procedure, number of expositions, average glandular dose, and complications.

METHODS:

The institutional review board approved this retrospective study and informed consent was waived. From July 2015 to January 2017, 306 patients with 306 suspicious mammographic findings (BI-RADS ⩾4) underwent mammography-guided biopsy, prone sVAB, or prone DBT-VAB. Student t test, chi-square, and multivariate regression statistics were used.

RESULTS:

During the study period, 155 prone sVAB procedures in 155 patients (mean age, 56 years; age range, 39-84 years) and 151 DBT-VABs in 151 patients (mean age, 57 years; age range, 33-84 years) were performed. Mean procedure time was shorter with DBT-VAB versus sVAB (14.5 versus 17.4 minutes, respectively; p < 0.001), and fewer images were acquired with DBT-VAB versus sVAB (8 vs 11, respectively; p < 0.001); the average glandular dose was significantly lower in DBT-VAB versus sVAB (11.8 mGy versus 18 mGy, respectively; p < 0.001). There were no differences in the distribution of histologic results (p = 0.74) or breast density (p = 0.09) between the two groups. No major complications were observed in either group.

CONCLUSION:

Performance of prone DBT-VAB was superior to prone sVAB because it allowed a faster procedure with fewer radiologic expositions and lower radiation dose.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Neoplasias de la Mama / Biopsia Guiada por Imagen Tipo de estudio: Diagnostic_studies / Observational_studies Límite: Adult / Aged / Aged80 / Female / Humans / Middle aged Idioma: En Revista: Tumori Año: 2022 Tipo del documento: Article País de afiliación: Italia Pais de publicación: EEUU / ESTADOS UNIDOS / ESTADOS UNIDOS DA AMERICA / EUA / UNITED STATES / UNITED STATES OF AMERICA / US / USA

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Neoplasias de la Mama / Biopsia Guiada por Imagen Tipo de estudio: Diagnostic_studies / Observational_studies Límite: Adult / Aged / Aged80 / Female / Humans / Middle aged Idioma: En Revista: Tumori Año: 2022 Tipo del documento: Article País de afiliación: Italia Pais de publicación: EEUU / ESTADOS UNIDOS / ESTADOS UNIDOS DA AMERICA / EUA / UNITED STATES / UNITED STATES OF AMERICA / US / USA