Skin dose measurement and estimating the dosimetric effect of applicator misplacement in gynecological brachytherapy: A patient and phantom study.
J Xray Sci Technol
; 29(5): 917-929, 2021.
Article
en En
| MEDLINE
| ID: mdl-34180462
OBJECTIVES: To evaluate skin dose differences between TPS (treatment planning system) calculations and TLD (thermo-luminescent dosimeters) measurements along with the dosimetric effect of applicator misplacement for patients diagnosed with gynecological (GYN) cancers undergoing brachytherapy. METHODS: The skin doses were measured using TLDs attached in different locations on patients' skin in pelvic regions (anterior, left, and right) for 20 patients, as well as on a phantom. In addition, the applicator surface dose was calculated with TLDs attached to the applicator. The measured doses were compared with TPS calculations to find TPS accuracy. For the phantom, different applicator shifts were applied to find the effect of applicator misplacement on the surface dose. RESULTS: The mean absolute dose differences between the TPS and TLDs results for anterior, left, and right points were 3.14±1.03, 6.25±1.88, and 6.20±1.97 %, respectively. The mean difference on the applicator surface was obtained 1.92±0.46 %. Applicator misplacements of 0.5, 2, and 4âcm (average of three locations) resulted in 9, 36, and 61%, dose errors respectively. CONCLUSIONS: The surface/skin differences between the calculations and measurements are higher in the left and right regions, which relate to the higher uncertainty of TPS dose calculation in these regions. Furthermore, applicator misplacements can result in high skin dose variations, therefore it can be an appropriate quality assurance method for future research.
Palabras clave
Texto completo:
1
Colección:
01-internacional
Base de datos:
MEDLINE
Asunto principal:
Braquiterapia
Límite:
Humans
Idioma:
En
Revista:
J Xray Sci Technol
Asunto de la revista:
RADIOLOGIA
Año:
2021
Tipo del documento:
Article
País de afiliación:
Irán
Pais de publicación:
Países Bajos