Your browser doesn't support javascript.
loading
Single-site 123I-FP-CIT reference values from individuals with non-degenerative parkinsonism-comparison with values from healthy volunteers.
Fahmi, Rachid; Platsch, Günther; Sadr, Alexandre Bani; Gouttard, Sylvain; Thobois, Stephane; Zuehlsdorff, Sven; Scheiber, Christian.
Afiliación
  • Fahmi R; Siemens Medical Solutions USA, Inc., Molecular Imaging, Knoxville, TN, USA.
  • Platsch G; Siemens Healthineers, Erlangen, Germany.
  • Sadr AB; Nuclear Medicine, Hospices Civils de Lyon, 69500, Bron, France.
  • Gouttard S; Nuclear Medicine, Hospices Civils de Lyon, 69500, Bron, France.
  • Thobois S; Movement Disorder Clinic, Pierre Wertheimer Neurologic Hospital, Hospices Civils de Lyon, 69500, Bron, France.
  • Zuehlsdorff S; Faculté de Médecine Lyon Sud, Université Claude Bernard Lyon 1, Lyon, France.
  • Scheiber C; Institut des Sciences Cognitives Marc Jeannerod, UMR 5229, CNRS, Bron, France.
Eur J Hybrid Imaging ; 4(1): 5, 2020 Mar 13.
Article en En | MEDLINE | ID: mdl-34191214
ABSTRACT

PURPOSE:

Iodine 123-radiolabeled 2ß-carbomethoxy-3ß-(4-iodophenyl)-N-(3-fluoropropyl) nortropane (123I-FP-CIT) SPECT can be performed to distinguish degenerative forms of movement disorders/parkinsonism/tremor from other entities such as idiopathic tremor or drug-induced parkinsonism. For equivocal cases, semi-quantification and comparison to reference values are a necessary addition to visual interpretation of 123I-FP-CIT scans. To overcome the challenges of multi-center recruitment and scanning of healthy volunteers, we generated 123I-FP-CIT reference values from individuals with various neurological conditions but without dopaminergic degeneration, scanned at a single center on the same SPECT-CT system following the same protocol, and compared them to references from a multi-center database built using healthy volunteers' data.

METHODS:

From a cohort of 1884 patients, we identified 237 subjects (120 men, 117 women, age range 16-88 years) through a two-stage selection process. Every patient had a final clinical diagnosis after a mean follow-up of 4.8 ± 1.3 years. Images were reconstructed using (1) Flash3D with scatter and CT-based attenuation corrections (AC) and (2) filtered back projection with Chang AC. Volume-of-interest analysis was performed using a commercial software to calculate specific binding ratios (SBRs), caudate-to-putamen ratios, and asymmetry values on different striatal regions. Generated reference values were assessed according to age and gender and compared with those from the ENC-DAT study, and their robustness was tested against a cohort of patients with different diagnoses.

RESULTS:

Age had a significant negative linear effect on all SBRs. Overall, the reduction rate per decade in SBR was between 3.80 and 5.70%. Women had greater SBRs than men, but this gender difference was only statistically significant for the Flash3D database. Linear regression was used to correct for age-dependency of SBRs and to allow comparisons to age-matched reference values and "normality" limits. Generated regression parameters and their 95% confidence intervals (CIs) were comparable to corresponding European Normal Control Database of DaTscan (ENC-DAT) results. For example, 95% CI mean slope for the striatum in women is - 0.015 ([- 0.019, - 0.011]) for the Flash3D database versus - 0.015 ([- 0.021, - 0.009]) for ENC-DAT. Caudate-to-putamen ratios and asymmetries were not influenced by age or gender.

CONCLUSION:

The generated 123I-FP-CIT references values have similar age-related distribution, with no increase in variance due to comorbidities when compared to values from a multi-center study with healthy volunteers. This makes it possible for sites to build their 123I-FP-CIT references from scans acquired during routine clinical practice.