Your browser doesn't support javascript.
loading
Air Flow Cooling Improves Anterolateral Thigh Perforator Mapping Using the FLIR ONE Thermal Camera.
Illg, Claudius; Krauss, Sabrina; Rothenberger, Jens; Kolbenschlag, Jonas; Daigeler, Adrien; Schäfer, Ruth Christine.
Afiliación
  • Illg C; Department of Hand, Plastic and Reconstructive Surgery, BG Unfallklinik Tuebingen, Eberhard Karls University Tuebingen, Germany.
  • Krauss S; Department of Hand, Plastic and Reconstructive Surgery, BG Unfallklinik Tuebingen, Eberhard Karls University Tuebingen, Germany.
  • Rothenberger J; Department of Plastic, Aesthetic, Reconstructive and Hand Surgery, Agaplesion Markus Hospital, Frankfurt, Germany.
  • Kolbenschlag J; Department of Hand, Plastic and Reconstructive Surgery, BG Unfallklinik Tuebingen, Eberhard Karls University Tuebingen, Germany.
  • Daigeler A; Department of Hand, Plastic and Reconstructive Surgery, BG Unfallklinik Tuebingen, Eberhard Karls University Tuebingen, Germany.
  • Schäfer RC; Department of Hand, Plastic and Reconstructive Surgery, BG Unfallklinik Tuebingen, Eberhard Karls University Tuebingen, Germany.
J Reconstr Microsurg ; 38(2): 144-150, 2022 Feb.
Article en En | MEDLINE | ID: mdl-34229351
ABSTRACT

BACKGROUND:

Due to the variable vascular anatomy preoperative perforator mapping facilitates anterolateral thigh (ALT) free flap harvesting. Dynamic infrared perforator imaging can assist preoperative planning by displaying hot spots that represent angiosomes. This study aims to compare previously described precooling methods to develop a standardized simplified protocol for ALT perforator planning.

METHODS:

Fifty thighs were examined with a FLIR ONE thermal camera. Four different cold challenges, including alcoholic disinfection, wet laparotomy sponge cooling, fan cooling, and cold pack application, were compared. Hot spot locations within a 250 mm × 80 mm area were compared double-blinded to perforator locations determined by Doppler ultrasonography considered as gold standard.

RESULTS:

The matching rate of thermographic hot spots and sonographically identified perforators was 34.9 ± 22.2%. An increased matching rate of 62.2 ± 42.2% was noted taking only favored perforators (septocutaneous course, diameter >1 mm, distance <3 cm to the center, and visible concomitant veins) into account. Precooling with a fan followed by alcoholic disinfection provided clearest thermograms and fastest results.

CONCLUSION:

Thermographic imaging is a reliable method for perforator imaging. Its supplemental use to ultrasound may reduce examination time and yield additional information. Precooling by air flow or alcoholic disinfection can be easily implemented and provide the best thermograms. The matching rate of thermographic hot spots and perforators increases when taking only clinically relevant perforators into account. Thermal perforator mapping therefore reduces distraction by negligible perforators.
Asunto(s)

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Procedimientos de Cirugía Plástica / Colgajos Tisulares Libres / Colgajo Perforante Tipo de estudio: Guideline / Prognostic_studies Idioma: En Revista: J Reconstr Microsurg Asunto de la revista: NEUROCIRURGIA Año: 2022 Tipo del documento: Article País de afiliación: Alemania

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Procedimientos de Cirugía Plástica / Colgajos Tisulares Libres / Colgajo Perforante Tipo de estudio: Guideline / Prognostic_studies Idioma: En Revista: J Reconstr Microsurg Asunto de la revista: NEUROCIRURGIA Año: 2022 Tipo del documento: Article País de afiliación: Alemania