Your browser doesn't support javascript.
loading
The wire loop technique for IVC filter removal.
Ochoa Chaar, Cassius Iyad; Kostiuk, Valentyna; Gholitabar, Navid.
Afiliación
  • Ochoa Chaar CI; Division of Vascular Surgery, Department of Surgery, Yale University School of Medicine, New Haven, Conn.
  • Kostiuk V; Yale University School of Medicine, New Haven, Conn.
  • Gholitabar N; Division of Vascular Surgery, Department of Surgery, Yale University School of Medicine, New Haven, Conn.
J Vasc Surg Cases Innov Tech ; 7(3): 369-370, 2021 Sep.
Article en En | MEDLINE | ID: mdl-34278059
ABSTRACT
Inferior vena cava (IVC) filters are effective therapy to prevent pulmonary embolism in patients with contraindication to anticoagulation. However, IVC wall penetration by the filter struts is a common complication that can lead to symptoms specially when adjacent organs are impacted. This case report and video describe the wire loop technique for successful endovascular IVC filter retrieval in a patient with lower back pain caused by a spinal strut penetration. The patient's back pain resolved after filter retrieval and he remained stable on anticoagulation with no recurrence of venous thromboembolism.
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Idioma: En Revista: J Vasc Surg Cases Innov Tech Año: 2021 Tipo del documento: Article

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Idioma: En Revista: J Vasc Surg Cases Innov Tech Año: 2021 Tipo del documento: Article
...