Your browser doesn't support javascript.
loading
Determination of serum carbamazepine and its metabolite by validated tandem mass spectrometric method and the effect of carbamazepine on various hematological and biochemical parameters.
Eryavuz Onmaz, Duygu; Abusoglu, Sedat; Ozturk, Bahadir; Abusoglu, Gulsum; Yerlikaya, Fatma Humeyra; Unlu, Ali.
Afiliación
  • Eryavuz Onmaz D; Department of Biochemistry, Selcuk University Faculty of Medicine, Konya, Turkey. Electronic address: duygu_eryavuz@hotmail.com.
  • Abusoglu S; Department of Biochemistry, Selcuk University Faculty of Medicine, Konya, Turkey.
  • Ozturk B; Department of Biochemistry, Selcuk University Faculty of Medicine, Konya, Turkey.
  • Abusoglu G; Department of Medical Laboratory Techniques, Selcuk University Vocational School of Health, Konya, Turkey.
  • Yerlikaya FH; Department of Biochemistry, Selcuk University Faculty of Medicine, Konya, Turkey.
  • Unlu A; Department of Biochemistry, Selcuk University Faculty of Medicine, Konya, Turkey.
J Pharm Biomed Anal ; 205: 114299, 2021 Oct 25.
Article en En | MEDLINE | ID: mdl-34388672
ABSTRACT

OBJECTIVES:

The study aimed to develop a validated LC-MS / MS method for the measurement of carbamazepine and carbamazepine-10,11-epoxide (CBZE) levels, to compare the carbamazepine concentrations measured by AbSciex API 3200 LC-MS/MS and Beckman Coulter Emit® 2000 immunoassay and to investigate the effect of carbamazepine concentrations on various hematological and biochemical parameters.

METHODS:

For the measurement of carbamazepine and CBZE levels, a validated LC-MS / MS method was developed. Serum carbamazepine levels of patients were measured by AbSciex API 3200 LC-MS / MS and Beckman Coulter Emit® 2000 immunoassay. Biochemical, hematological parameters, and hormone levels were measured by Beckman-Coulter AU 5800 (Beckman Coulter, Brea, USA), Beckman Coulter LH 780 (Beckman Coulter, Miami, FL, USA), and Cobas 6000 (Roche Diagnostics, Germany) analyzers, respectively.

RESULTS:

The imprecision values for all analytes were less than 7.0 %. The correlation coefficient between the methods was 0.981 (95 % confidence interval 0.975 to 0.985). Linear regression analysis demonstrated highly significant associations of carbamazepine concentrations with albumin (B = -0.300, p = 0.040), calcium (B = -0.262, p = 0.004), phosphorus (B = -0.241, p = 0.022), WBC (B = -0.288, p = <0.001), PLT (B = -0.236, p = 0.003), RBC (B = -0.257, p = 0.001), NEU% (B = -0.289, p = <0.001), LYM% (B = -0.268, p = 0.001), D vitamini (B = -0.344, p = 0.006) levels.

CONCLUSIONS:

A robust, rapid, and simple method has been developed. Our study revealed that carbamazepine and its metabolite have a significant correlation and likely impact on bone metabolism, blood cell counts, serum protein, albumin levels, and electrolyte concentrations.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Carbamazepina / Espectrometría de Masas en Tándem Límite: Humans Idioma: En Revista: J Pharm Biomed Anal Año: 2021 Tipo del documento: Article

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Carbamazepina / Espectrometría de Masas en Tándem Límite: Humans Idioma: En Revista: J Pharm Biomed Anal Año: 2021 Tipo del documento: Article