Feasibility of operating a millimeter-scale graphite calorimeter for absolute dosimetry of small-field photon beams in the clinic.
Côté, Benjamin; Keszti, Federico; Bancheri, Julien; Sarfehnia, Arman; Seuntjens, Jan; Renaud, James.
Afiliación
  • Côté B; Medical Physics Unit, Gerald Bronfman Department of Oncology, McGill University, Montreal, Quebec, Canada.
  • Keszti F; Medical Physics Unit, Gerald Bronfman Department of Oncology, McGill University, Montreal, Quebec, Canada.
  • Bancheri J; Medical Physics Unit, Gerald Bronfman Department of Oncology, McGill University, Montreal, Quebec, Canada.
  • Sarfehnia A; Medical Physics Unit, Gerald Bronfman Department of Oncology, McGill University, Montreal, Quebec, Canada.
  • Seuntjens J; Department of Radiation Oncology, University of Toronto, Toronto, Ontario, Canada.
  • Renaud J; Medical Physics Unit, Gerald Bronfman Department of Oncology, McGill University, Montreal, Quebec, Canada.
Med Phys ; 48(11): 7476-7492, 2021 Nov.
Article en En | MEDLINE | ID: mdl-34549805
ABSTRACT

PURPOSE:

To characterize and build a cylindrically layered graphite calorimeter the size of a thimble ionization chamber for absolute dosimetry of small fields. This detector has been designed in a familiar probe format to facilitate integration into the clinical workflow. The feasibility of operating this absorbed dose calorimeter in quasi-adiabatic mode is assessed for high-energy accelerator-based photon beams.

METHODS:

This detector, herein referred to as Aerrow MK7, is a miniaturized version of a previously validated aerogel-insulated graphite calorimeter known as Aerrow. The new model was designed and developed using numerical methods. Medium conversion factors from graphite to water, small-field output correction factors, and layer perturbation factors for this dosimeter were calculated using the EGSnrc Monte Carlo code system. A range of commercially available aerogel densities were studied for the insulating layers, and an optimal density was selected by minimizing the small-field output correction factors. Heat exchange within the detector was simulated using a five-body compartmental heat transfer model. In quasi-adiabatic mode, the sensitive volume (a 3 mm diameter cylindrical graphite core) experiences a temperature rise during irradiation on the order of 1.3 mK·Gy-1 . The absorbed dose is obtained by calculating the product of this temperature rise with the specific heat capacity of the graphite. The detector was irradiated with 6 MV ( % dd ( 10 ) x  = 63.5%) and 10 MV ( % dd ( 10 ) x  = 71.1%) flattening filter-free (FFF) photon beams for two field sizes, characterized by S clin dimensions of 2.16 and 11.0 cm. The dose readings were compared against a calibrated Exradin A1SL ionization chamber. All dose values are reported at d max in water.

RESULTS:

The field output correction factors for this dosimeter design were computed for field sizes ranging from S clin  = 0.54 to 11.0 cm. For all aerogel densities studied, these correction factors did not exceed 1.5%. The relative dose difference between the two dosimeters ranged between 0.3% and 0.7% for all beams and field sizes. The smallest field size experimentally investigated, S clin  = 2.16 cm, which was irradiated with the 10 MV FFF beam, produced readings of 84.4 cGy (±1.3%) in the calorimeter and 84.5 cGy (±1.3%) in the ionization chamber.

CONCLUSION:

The median relative difference in absorbed dose values between a calibrated A1SL ionization chamber and the proposed novel graphite calorimeter was 0.6%. This preliminary experimental validation demonstrates that Aerrow MK7 is capable of accurate and reproducible absorbed dose measurements in quasi-adiabatic mode.