DOSIMETRIC EVALUATION OF THE TWO-PHASE COMPUTED TOMOGRAPHY IN PARATHYROID GLANDS IMAGING.
Dimitroukas, Christos P; Metaxas, Vasileios I; Efthymiou, Fotios O; Zampakis, Petros E; Kalogeropoulou, Christina P; Panayiotakis, George S.
Afiliación
  • Dimitroukas CP; Department of Medical Physics, School of Medicine, University of Patras, Patras 265 04, Greece.
  • Metaxas VI; Department of Medical Physics, University Hospital of Patras, Patras 265 04, Greece.
  • Efthymiou FO; Department of Medical Physics, School of Medicine, University of Patras, Patras 265 04, Greece.
  • Zampakis PE; Department of Medical Physics, School of Medicine, University of Patras, Patras 265 04, Greece.
  • Kalogeropoulou CP; Department of Radiology, School of Medicine, University of Patras, Patras 265 04, Greece.
  • Panayiotakis GS; Department of Radiology, University Hospital of Patras, Patras 265 04, Greece.
Radiat Prot Dosimetry ; 196(3-4): 207-219, 2021 Nov 12.
Article en En | MEDLINE | ID: mdl-34635914
This study evaluates the patient radiation dose from the two-phase protocols of two different computed tomography (CT) systems and compares this with that delivered by the other similar protocols previously published. Two hundred and fourteen patients with primary hyperparathyroidism were included in the study with a two-phase CT scan between 2008 and 2020 by using a Toshiba Aquilion Prime 80 and a GE Light Speed 16. The standard 'neck' or a modified 'parathyroid' protocol was used. The patient dose was evaluated in terms of volumetric computed tomography dose index (CTDIvol), dose length product (DLP) and effective dose (ED) per acquisition protocol and CT system. CTDIvol and DLP were recorded retrospectively, while the ED was calculated based on DLP and an appropriate conversion coefficient. Comparisons of patient dose between the two protocols and two CT systems and the corresponding published values were established. A significantly lower patient dose (40.2-43.2%) than the GE system (p < 0.0001) resulted from the Toshiba system. The 'parathyroid' protocol resulted in a 6.5-9.6% lower patient dose than the standard 'neck' protocol. Compared with the literature, the lowest ED value (3.6 mSv) was observed since this protocol consists of a lowered tube voltage of 100 kVp, a reduced scan length for the pre-contrast phase and implementation of an iterative reconstruction algorithm.
Asunto(s)

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Glándulas Paratiroides / Tomografía Computarizada por Rayos X Tipo de estudio: Observational_studies Límite: Humans Idioma: En Revista: Radiat Prot Dosimetry Año: 2021 Tipo del documento: Article País de afiliación: Grecia Pais de publicación: Reino Unido

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Glándulas Paratiroides / Tomografía Computarizada por Rayos X Tipo de estudio: Observational_studies Límite: Humans Idioma: En Revista: Radiat Prot Dosimetry Año: 2021 Tipo del documento: Article País de afiliación: Grecia Pais de publicación: Reino Unido