Usefulness of intraoperative C-arm image intensifier in reducing errors of acetabular component during primary total hip arthroplasty: an application of Widmer's method.
BMC Musculoskelet Disord
; 22(1): 892, 2021 Oct 20.
Article
en En
| MEDLINE
| ID: mdl-34670523
BACKGROUND: Acetabular prosthesis positioning in total hip arthroplasty (THA) is crucial in reducing the risk of dislocation. There has been minimal research on the proper way to put the acetabular components into the safe zone intraoperatively. Assessment of version by intraoperative imaging intensifier is very valuable. The value of Widmer's method, using the intraoperative C-arm available to determine cup anteversion was assessed. METHODS: One hundred one hips in 91 patients who underwent primary THA were eligible for inclusion. Utilizing intraoperative C-arm images, measurement was performed using the technique described by Widmer. The values obtained using 3D computed tomography postoperatively, which determined the anteversion of the acetabular component, were regarded as the reference standard. RESULTS: The method of Widmer obtained values similar to those obtained using 3D computed tomography and was considered accurate (n.s.). All 101 hips were positioned in the set target zone. Among the 101 hips, the cup position in nine hips (8.9%) was changed. The dislocation rate in our study was 1.0% with all dislocations occurring in hips placed in the target zone. The mean Harris hip score after THA in 1 year was 94.2 (82-98). CONCLUSIONS: The method of Widmer was accurate using intraoperative imaging intensifier for the measurement of the anteversion of the acetabular component during THA, with reference to the anteversion obtained from the 3D computed tomography. Also, utilizing intraoperative C-arm imaging was very useful because it allowed for correction of the position of the acetabular cup.
Palabras clave
Texto completo:
1
Colección:
01-internacional
Base de datos:
MEDLINE
Asunto principal:
Artroplastia de Reemplazo de Cadera
/
Luxaciones Articulares
/
Prótesis de Cadera
Límite:
Humans
Idioma:
En
Revista:
BMC Musculoskelet Disord
Asunto de la revista:
FISIOLOGIA
/
ORTOPEDIA
Año:
2021
Tipo del documento:
Article
Pais de publicación:
Reino Unido