Influence of age and scanning system on the learning curve of experienced and novel intraoral scanner operators: A multi-centric clinical trial.
Zarauz, Cristina; Sailer, Irena; Pitta, João; Robles-Medina, Mercedes; Hussein, Abra Abdulahai; Pradíes, Guillermo.
Afiliación
  • Zarauz C; Division of Fixed Prosthodontics and Biomaterials, Clinic of Dental Medicine, University of Geneva, Rue Michel-Servet 1, 1211 Genève 4, Switzerland. Electronic address: cristina.zarauz@unige.ch.
  • Sailer I; Division of Fixed Prosthodontics and Biomaterials, Clinic of Dental Medicine, University of Geneva, Rue Michel-Servet 1, 1211 Genève 4, Switzerland.
  • Pitta J; Division of Fixed Prosthodontics and Biomaterials, Clinic of Dental Medicine, University of Geneva, Rue Michel-Servet 1, 1211 Genève 4, Switzerland.
  • Robles-Medina M; Department of Prosthodontics, Faculty of Dentistry, Complutense University of Madrid, Pza. Ramón y Cajal, s/n, Madrid 28040, Spain.
  • Hussein AA; Division of Fixed Prosthodontics and Biomaterials, Clinic of Dental Medicine, University of Geneva, Rue Michel-Servet 1, 1211 Genève 4, Switzerland.
  • Pradíes G; Department of Prosthodontics, Faculty of Dentistry, Complutense University of Madrid, Pza. Ramón y Cajal, s/n, Madrid 28040, Spain.
J Dent ; 115: 103860, 2021 12.
Article en En | MEDLINE | ID: mdl-34715248
ABSTRACT

OBJECTIVES:

To evaluate the effect of age and intra-oral scanner (IOS) on the learning curve of inexperienced operators.

METHODS:

Thirty-four operators pertaining to 1 of 3 groups (G1) students ≤ 25 years (y), (G2) dentists ≥ 40y, and (G3) a control group of experienced IOS operators (no age limitation), were included. All participants performed baseline and final quadrant scans on a volunteer subject, before and after a training program of 3 sessions, with two different IOS TRIOS 3 (S1) and True Definition (S2). Baseline and final scanning times were registered in seconds. A Pearson correlation was applied to evaluate the correlation between age and scanning time. An ANOVA of repeated measures test was applied to evaluate inter-group (G1, G2, G3) and inter-system performance. Significance level was set at a = 0.05.

RESULTS:

Age and scanning time for inexperienced operators showed a weak positive correlation for final scanning time (r = 0.29, p < 0.05). When comparing groups and filtering by IOS, S1 failed to show differences between groups (p > 0.05). With S2, the control group demonstrated a better performance than G2 (p < 0.05), while G1 only demonstrated a better performance than G2 at final scanning time (p = 0.005). Overall, the type of IOS had a significant impact on the scanning time (p < 0.001).

CONCLUSION:

Results from this study indicate that age and type of IOS have an impact on the performance and learning curve of inexperienced IOS operators. CLINICAL

SIGNIFICANCE:

Gaining knowledge on how different aspects, such as age, experience or IOS system, influence the learning curve to IOSs is relevant due to the financial and strategical impact associated with the acquisition of an IOS.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Técnica de Impresión Dental / Curva de Aprendizaje Tipo de estudio: Clinical_trials Límite: Humans Idioma: En Revista: J Dent Año: 2021 Tipo del documento: Article

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Técnica de Impresión Dental / Curva de Aprendizaje Tipo de estudio: Clinical_trials Límite: Humans Idioma: En Revista: J Dent Año: 2021 Tipo del documento: Article