Simplifying the hardware requirements for fast neural EIT of peripheral nerves.
Ravagli, Enrico; Mastitskaya, Svetlana; Holder, David; Aristovich, Kirill.
Afiliación
  • Ravagli E; Medical Physics and Biomedical Engineering, University College London, United Kingdom.
  • Mastitskaya S; Medical Physics and Biomedical Engineering, University College London, United Kingdom.
  • Holder D; Medical Physics and Biomedical Engineering, University College London, United Kingdom.
  • Aristovich K; Medical Physics and Biomedical Engineering, University College London, United Kingdom.
Physiol Meas ; 43(1)2022 01 31.
Article en En | MEDLINE | ID: mdl-34915462
ABSTRACT
Objective. The main objective of this study was to assess the feasibility of lowering the hardware requirements for fast neural electrical impedance tomography (EIT) in order to support the distribution of this technique. Specifically, the feasibility of replacing the commercial modules present in the existing high-end setup with compact and cheap customized circuitry was assessed.Approach. Nerve EIT imaging was performed on rat sciatic nerves with both our standard ScouseTom setup and a customized version in which commercial benchtop current sources were replaced by custom circuitry. Electrophysiological data and images collected in the same experimental conditions with the two setups were compared. Data from the customized setup was subject to a down-sampling analysis to simulate the use of a recording module with lower specifications.Main results. Compound action potentials (573 ± 287µV and 487 ± 279µV,p=0.28) and impedance changes (36 ± 14µV and 31 ± 16µV,p=0.49) did not differ significantly when measured using commercial high-end current sources or our custom circuitry, respectively. Images reconstructed from both setups showed neglibile (<1voxel, i.e. 40µm) difference in peak location and a high degree of correlation (R2 = 0.97). When down-sampling from 24 to 16 bits ADC resolution and from 100 to 50 KHz sampling frequency, signal-to-noise ratio showed acceptable decrease (<-20%), and no meaningful image quality loss was detected (peak location difference <1voxel, pixel-by-pixel correlationR2 = 0.99).

Significance:

The technology developed for this study greatly reduces the cost and size of a fast neural EIT setup without impacting quality and thus promotes the adoption of this technique by the neuroscience research community.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Nervio Ciático / Tomografía Límite: Animals Idioma: En Revista: Physiol Meas Asunto de la revista: BIOFISICA / ENGENHARIA BIOMEDICA / FISIOLOGIA Año: 2022 Tipo del documento: Article País de afiliación: Reino Unido

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Nervio Ciático / Tomografía Límite: Animals Idioma: En Revista: Physiol Meas Asunto de la revista: BIOFISICA / ENGENHARIA BIOMEDICA / FISIOLOGIA Año: 2022 Tipo del documento: Article País de afiliación: Reino Unido