Your browser doesn't support javascript.
loading
Comparison of the performance of the CareStart Malaria Pf/Pan Combo test and field microscopy in the diagnosis of Plasmodium vivax malaria in North Sumatera, Indonesia.
Pasaribu, Ayodhia Pitaloka; Nasution, Irma Sari; Sembiring, Krisnarta; Fahmi, Fahmi; Pasaribu, Syahril.
Afiliación
  • Pasaribu AP; Department of Child Health, Medical Faculty, Universitas Sumatera Utara, Medan, 20155, Indonesia. ayodhiapp@usu.ac.id.
  • Nasution IS; Medical Faculty, Universitas Sumatera Utara, Medan, 20155, Indonesia.
  • Sembiring K; Department of Child Health, Medical Faculty, Universitas Sumatera Utara, Medan, 20155, Indonesia.
  • Fahmi F; Department of Electrical Engineering, Universitas Sumatera Utara, Medan, 20155, Indonesia.
  • Pasaribu S; Department of Child Health, Medical Faculty, Universitas Sumatera Utara, Medan, 20155, Indonesia.
Malar J ; 21(1): 27, 2022 Jan 29.
Article en En | MEDLINE | ID: mdl-35093067
ABSTRACT

BACKGROUND:

In areas where malaria is endemic and where trained microscopists are not available, rapid diagnostic tests (RDTs) are needed not only to allow prompt treatment without delay but also to prevent overdiagnosis and overtreatment based on clinical judgements that may lead to drug resistance. This study aimed to compare the performances of the CareStart Pf/Pan Combo test to field microscopy, which is considered to be the gold standard for malaria diagnosis.

METHODS:

Any person with a fever or a history of fever within 48 h who came to the health centre was recruited for the study and tested both by the CareStart Pf/Pan test and by field microscopy. Sensitivity, specificity, positive predictive value, negative predictive value, and accuracy were analysed with both methods.

RESULTS:

Two-hundred study participants were enrolled 96 (48%) were found to be positive through microscopy, while 100 (50%) participants were found to be positive through RDT. The RDT produced four false-positive results. High sensitivity and specificity were observed for the CareStart Pf/Pan test (100 and 96.15%, respectively). The CareStart Pf/Pan test also showed excellent agreement with the field microscopy results.

CONCLUSION:

The Carestart Pf/Pan could be used as an alternative diagnostic test in malaria-endemic areas where facility for performing microscopy is not available.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Malaria Vivax / Pruebas Diagnósticas de Rutina / Microscopía Tipo de estudio: Diagnostic_studies / Prognostic_studies Límite: Humans País/Región como asunto: Asia Idioma: En Revista: Malar J Asunto de la revista: MEDICINA TROPICAL Año: 2022 Tipo del documento: Article País de afiliación: Indonesia

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Malaria Vivax / Pruebas Diagnósticas de Rutina / Microscopía Tipo de estudio: Diagnostic_studies / Prognostic_studies Límite: Humans País/Región como asunto: Asia Idioma: En Revista: Malar J Asunto de la revista: MEDICINA TROPICAL Año: 2022 Tipo del documento: Article País de afiliación: Indonesia