Validity and Reliability of the Insole3 Instrumented Shoe Insole for Ground Reaction Force Measurement during Walking and Running.
Cramer, Leora A; Wimmer, Markus A; Malloy, Philip; O'Keefe, Joan A; Knowlton, Christopher B; Ferrigno, Christopher.
Afiliación
  • Cramer LA; Department of Orthopedic Surgery, Rush University Medical Center, 1611 W Harrison Street, Suite 201, Chicago, IL 60612, USA.
  • Wimmer MA; Department of Orthopedic Surgery, Rush University Medical Center, 1611 W Harrison Street, Suite 201, Chicago, IL 60612, USA.
  • Malloy P; Department of Anatomy and Cell Biology, Rush University Medical Center, 600 S. Paulina St., Suite 507, Chicago, IL 60612, USA.
  • O'Keefe JA; Department of Orthopedic Surgery, Rush University Medical Center, 1611 W Harrison Street, Suite 201, Chicago, IL 60612, USA.
  • Knowlton CB; Department of Physical Therapy, Arcadia University, 450 S. Easton Road, Glenside, PA 19038, USA.
  • Ferrigno C; Department of Anatomy and Cell Biology, Rush University Medical Center, 600 S. Paulina St., Suite 507, Chicago, IL 60612, USA.
Sensors (Basel) ; 22(6)2022 Mar 11.
Article en En | MEDLINE | ID: mdl-35336374
ABSTRACT
Pressure-detecting insoles such as the Insole3 have potential as a portable alternative for assessing vertical ground reaction force (vGRF) outside of specialized laboratories. This study evaluated whether the Insole3 is a valid and reliable alternative to force plates for measuring vGRF. Eleven healthy participants walked overground at slow and moderately paced speeds and ran at a moderate pace while collecting vGRF simultaneously from a force plate (3000 Hz) and Insole3 (100 Hz). Intraclass correlation coefficients (ICC) demonstrated excellent vGRF agreement between systems during both walking speeds for Peak 1, Peak 2, the valley between peaks, and the vGRF impulse (ICC > 0.941). There was excellent agreement during running for the single vGRF peak (ICC = 0.942) and impulse (ICC = 0.940). The insoles slightly underestimated vGRF peaks (−3.7% to 0.9% bias) and valleys (−2.2% to −1.8% bias), and slightly overestimated impulses (4.2% to 5.6% bias). Reliability between visits for all three activities was excellent (ICC > 0.970). The Insole3 is a valid and reliable alternative to traditional force plates for assessing vGRF during walking and running in healthy adults. The excellent ICC values during slow walking suggests that the Insole3 may be particularly suitable for older adults in clinical and home settings.