Your browser doesn't support javascript.
loading
Median 10-year whole blood metal ion levels and clinical outcome of ReCap-M2a-Magnum metal-on-metal total hip arthroplasty.
Pietiläinen, Sakari; Linnovaara, Axel; Venäläinen, Mikko S; Mäntymäki, Heikki; Laaksonen, Inari; Lankinen, Petteri; Mäkelä, Keijo T.
Afiliación
  • Pietiläinen S; Department of Orthopaedics and Traumatology, Turku University Hospital; University of Turku; Department of Orthopaedics and Traumatology, Vaasa Central Hospital. sakari.pietilainen@gmail.com.
  • Linnovaara A; Department of Orthopaedics and Traumatology, Vaasa Central Hospital. axel.linnovaara@pp.inet.fi.
  • Venäläinen MS; Turku Bioscience Centre, University of Turku and Åbo Akademi University; Turku University Hospital, Division of Medical Imaging, Department of Medical Physics, Turku. mikko.venalainen@utu.fi.
  • Mäntymäki H; Department of Orthopaedics, Tampere University Hospital, Tampere. heikki.mantymaki@gmail.com.
  • Laaksonen I; Department of Orthopaedics and Traumatology, Turku University Hospital; University of Turku. Inari.laaksonen@tyks.fi.
  • Lankinen P; Department of Orthopaedics and Traumatology, Turku University Hospital; University of Turku; Satasairaala Central Hospital, Satakunta Hospital District, Pori, Finland. petteri.lankinen@utu.fi.
  • Mäkelä KT; Department of Orthopaedics and Traumatology, Turku University Hospital; University of Turku. keijo.makela@tyks.fi.
Acta Orthop ; 93: 444-450, 2022 04 19.
Article en En | MEDLINE | ID: mdl-35438184
ABSTRACT
BACKGROUND AND

PURPOSE:

We have previously reported that the whole blood (WB) chromium (Cr) and cobalt (Co) ion levels decrease in the short term after ReCap-M2a-Magnum large-diameter head (LDH) metal-on-metal (MoM) total hip arthroplasty (THA). This study reports long-term metal ion levels and clinical outcomes after ReCap-Magnum THA. PATIENTS AND

METHODS:

ReCap-M2a-Magnum LDH THA was used in 1,450 patients in our hospital district from 2005 to 2012. Median follow-up time was 10 years. 991 patients had 2 or more metal ion measurements. The median measurement interval was 4 years. Individual metal ion change was assessed using logarithmic metal ion values in a random coefficient model. Kaplan-Meier survival estimates were calculated for revision surgery for any reason for revision, and separately for metal-related adverse events (metal ions above safe upper limit [SUL], revision due to ARMD, or pseudotumor).

RESULTS:

Geometric mean of Cr decreased from 1.8 ppb (geometric standard deviation [GSD] 1.8) to 1.0 ppb (GSD 2.8, p < 0.001). The Co levels decreased from 1.7 ppb (GSD 2.4) to 1.4 ppb (GSD 2.8, p < 0.001). The hip-specific survival was 85% for revision due to any reason at 14 years and the hip-specific survival for any metal-related adverse event was 69% at 14 years.

INTERPRETATION:

WB Cr and Co levels continued to decrease in the long-term follow-up of ReCap-M2a-Magnum THA patients. The amount of metal-related adverse events was rather high, but revision surgery was seldom required. We suggest that after 10 years from the implantation a 5-year measurement interval may be sufficient for asymptomatic ReCap-M2a-Magnum patients.
Asunto(s)

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Artroplastia de Reemplazo de Cadera / Prótesis Articulares de Metal sobre Metal / Prótesis de Cadera Límite: Humans Idioma: En Revista: Acta Orthop Asunto de la revista: ORTOPEDIA Año: 2022 Tipo del documento: Article Pais de publicación: SE / SUECIA / SUÉCIA / SWEDEN

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Artroplastia de Reemplazo de Cadera / Prótesis Articulares de Metal sobre Metal / Prótesis de Cadera Límite: Humans Idioma: En Revista: Acta Orthop Asunto de la revista: ORTOPEDIA Año: 2022 Tipo del documento: Article Pais de publicación: SE / SUECIA / SUÉCIA / SWEDEN