Depth of cure of 10 resin-based composites light-activated using a laser diode, multi-peak, and single-peak light-emitting diode curing lights.
Rocha, Mateus Garcia; Maucoski, Cristiane; Roulet, Jean-François; Price, Richard Bengt.
Afiliación
  • Rocha MG; Department of Restorative Dental Science, Center of Dental Biomaterials College of Dentistry, University of Florida, Gainesville, FL, United States.
  • Maucoski C; Department of Dental Clinical Sciences, Faculty of Dentistry, Dalhousie University, 5981 University Ave., Halifax, Nova Scotia B3H 4R2, Canada; Department of Restorative Dentistry, State University of Ponta Grossa, Ponta Grossa, Parana, Brazil.
  • Roulet JF; Department of Restorative Dental Science, Center of Dental Biomaterials College of Dentistry, University of Florida, Gainesville, FL, United States.
  • Price RB; Department of Dental Clinical Sciences, Faculty of Dentistry, Dalhousie University, 5981 University Ave., Halifax, Nova Scotia B3H 4R2, Canada. Electronic address: rbprice@dal.ca.
J Dent ; 122: 104141, 2022 07.
Article en En | MEDLINE | ID: mdl-35483497
ABSTRACT

OBJECTIVES:

To evaluate the depth of cure (DOC) of ten contemporary resin-based composites (RBCs), light-cured using different LCUs and exposure times.

METHODS:

The power, radiant emittance, irradiance, radiant exposure (RE), and beam profiles from a laser (M, Monet), a multi-peak (V, Valo Grand), and single-peak (S, SmartLite Pro) LCU were measured. The DOC was measured using a 6-mm diameter metal mold and a solvent dissolution method to remove the uncured RBC. The length of the remaining RBC was divided by 2. The exposure times were 1 s and 3 s for M, 10 s and 20 s for V, and 10 s and 20 s for S. Data were analyzed using Bland-Altman distribution, Pearson's Correlation, and an artificial neural network (ANN) to establish the relative importance of the factors on the DOC (α=0.05; ß=0.2).

RESULTS:

Significant differences were found in the DOC achieved by the different LCUs and composites. The laser LCU emitted the highest power, radiant emittance is used above and in the tables and delivered the highest irradiance. However, this LCU used for 1 s delivered the lowest RE and produced the shortest DOC in all ten RBCs. The ANN demonstrated that the RE is the most critical factor for the DOC. Bland-Altman comparisons showed that the DOCs achieved with the laser LCU used for 1 s were between 17 and 34% shorter than the other conditions.

CONCLUSIONS:

Although the laser LCU cured all 10 RBCs when used for 1 s, it produced the shallowest DOC, and some RBCs did not achieve their minimum DOC threshold. The RE and not the irradiance was the most important factor in determining the DOC of these 10 RBCs. CLINICAL

SIGNIFICANCE:

Despite delivering high power and irradiance, the laser used for l s delivered a lower radiant exposure than the conventional LCUs used for 10 s. This resulted in a shorter DOC.