Your browser doesn't support javascript.
loading
Prohibitive Radial Artery Occlusion Rates Following Transradial Access Using a 6-French Neuron MAX Long Sheath for Intracranial Aneurysm Treatment.
Boeken, Tom; Altayeb, Adnan; Shotar, Eimad; Premat, Kévin; Lenck, Stéphanie; Boch, Anne-Laure; Drir, Mehdi; Sourour, Nader-Antoine; Clarençon, Frédéric.
Afiliación
  • Boeken T; Department of Neuroradiology, Sorbonne University. APHP. Pitié-Salpêtrière Hospital, 47, Bd de l'Hôpital, 75013, Paris, France.
  • Altayeb A; Department of Neuroradiology, Sorbonne University. APHP. Pitié-Salpêtrière Hospital, 47, Bd de l'Hôpital, 75013, Paris, France.
  • Shotar E; Department of Neuroradiology, Sorbonne University. APHP. Pitié-Salpêtrière Hospital, 47, Bd de l'Hôpital, 75013, Paris, France.
  • Premat K; Department of Neuroradiology, Sorbonne University. APHP. Pitié-Salpêtrière Hospital, 47, Bd de l'Hôpital, 75013, Paris, France.
  • Lenck S; Department of Neuroradiology, Sorbonne University. APHP. Pitié-Salpêtrière Hospital, 47, Bd de l'Hôpital, 75013, Paris, France.
  • Boch AL; Department of Neurosurgery, Sorbonne University. APHP. Pitié-Salpêtrière Hospital, Paris, France.
  • Drir M; Department of Anesthesiology, Sorbonne University. APHP. Pitié-Salpêtrière Hospital, Paris, France.
  • Sourour NA; Department of Neuroradiology, Sorbonne University. APHP. Pitié-Salpêtrière Hospital, 47, Bd de l'Hôpital, 75013, Paris, France.
  • Clarençon F; Department of Neuroradiology, Sorbonne University. APHP. Pitié-Salpêtrière Hospital, 47, Bd de l'Hôpital, 75013, Paris, France. frederic.clarencon@aphp.fr.
Clin Neuroradiol ; 32(4): 1031-1036, 2022 Dec.
Article en En | MEDLINE | ID: mdl-35551420
ABSTRACT

OBJECTIVE:

To evaluate the feasibility, safety, and rate of radial artery occlusion (RAO) after the treatment of intracranial aneurysms using a 6F Neuron MAX (Penumbra, Alameda, CA, USA) long sheath directly into the radial artery.

METHODS:

All consecutive patients treated for unruptured intracranial aneurysms with TRA using a 6F Neuron MAX catheter between September 2019 and May 2021 in a single tertiary center were screened. They were referred to a consultation and an ultrasound-Doppler assessment of the radial artery 3 months after treatment with the attending neuroradiologist. Patients with available assessment of the radial artery patency were included.

RESULTS:

A total of 17 patients (median age 58 years, range 35-68 years; sex ratio F/M 15/2) were treated for intracranial aneurysms using a 6F Neuron MAX directly into the right radial artery and included. Treatment was a technical success for 16/17 (94%) patients and 1 patient (6%) required a conversion to femoral access. The median radial artery diameter at the puncture site was 2.7 mm (range 1.8-2.9mm). No symptomatic RAO was noted during follow-up. Assessment at 3 months revealed 7/17 (41%) asymptomatic RAOs.

CONCLUSION:

Even if technically feasible, the use of a 6F Neuron Max long sheath for triaxial catheterization in intracranial interventions, especially flow diversion, may be responsible for a high radial artery occlusion rate (41%). Although being asymptomatic in all cases in our series, this high occlusion rate may be a concern for further interventions. The development of dedicated radial long sheaths for neurointerventions, with external hydrophilic coating, seems necessary.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Arteriopatías Oclusivas / Aneurisma Intracraneal Límite: Adult / Aged / Humans / Middle aged Idioma: En Revista: Clin Neuroradiol Asunto de la revista: NEUROLOGIA / RADIOLOGIA Año: 2022 Tipo del documento: Article País de afiliación: Francia

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Arteriopatías Oclusivas / Aneurisma Intracraneal Límite: Adult / Aged / Humans / Middle aged Idioma: En Revista: Clin Neuroradiol Asunto de la revista: NEUROLOGIA / RADIOLOGIA Año: 2022 Tipo del documento: Article País de afiliación: Francia