Your browser doesn't support javascript.
loading
Initial Airtightness of the Lung Parenchyma After Transection of the Interlobar Fissure - Monopolar Cutter, Stapler Versus Nd: YAG Laser.
Kirschbaum, Andreas; Eigendorf, Peter; Wiesmann, Thomas; Mirow, Nikolas; Meyer, Christian.
Afiliación
  • Kirschbaum A; Department of Visceral, Thoracic and Vascular Surgery, University Hospital Gießen and Marburg (UKGM), Marburg, Germany. Electronic address: Kirschbaum001@gmx.de.
  • Eigendorf P; Department of Surgery, Asklepios Stadtklinik Bad Wildungen, Bad Wildungen, Germany.
  • Wiesmann T; Department of Anaesthesiology, University Hospital Gießen and Marburg (UKGM), Marburg, Germany.
  • Mirow N; Department of Cardiac Surgery, University Hospital Gießen and Marburg (UKGM), Marburg, Germany.
  • Meyer C; Department of Surgery, Asklepios Stadtklinik Bad Wildungen, Bad Wildungen, Germany.
J Surg Res ; 278: 79-85, 2022 10.
Article en En | MEDLINE | ID: mdl-35594618
ABSTRACT

INTRODUCTION:

Every lobectomy requires the transection of the interlobar fissure. Resection surfaces must be airtight in order to avoid leakage and infection. Using an ex vivo model based on porcine lung, we compared three techniques with respect to initial airtightness at different inspiratory pressures. MATERIALS AND

METHODS:

In the first technique (group 1), we transected the interlobar fissure with a monopolar cutter and overstitched the edges of the resection area with a monofilament thread. In the second technique (group 2), the interlobar fissure was cut with a stapling device. In the third technique (group 3), the interlobar fissure was cut using a laser fibre connected to an Nd YAG laser. The resection areas were not overstitched; 15 transections were performed in each group.

RESULTS:

In group 1, three parenchymatous bridges leaked starting at a pressure of 25 mbar. In the other two groups, all preparations were airtight at this pressure. If the ventilation pressure was increased up to 40 mbar, all seams in group 1 were leaky at a pressure of 35 mbar. Four staple seams were airtight at a pressure of 40 mbar. In group 3, 11 preparations (73.3 %) were airtight up to a pressure of 40 mbar.

CONCLUSIONS:

Based on our results, the use of an Nd YAG laser is suitable for the transection of the interlobar fissure. In effect, this technique compares well with the other techniques examined.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Terapia por Láser / Láseres de Estado Sólido Límite: Animals Idioma: En Revista: J Surg Res Año: 2022 Tipo del documento: Article

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Terapia por Láser / Láseres de Estado Sólido Límite: Animals Idioma: En Revista: J Surg Res Año: 2022 Tipo del documento: Article