[Evaluation of Sysmex XN-3000 for Detection of Schistocyte].
Hou, Dan-Feng; He, Xue; Diao, Ying-Ying.
Afiliación
  • Hou DF; Laboratory Medicine Innovation Unit,Chinese Academy of Medical Sciences; Department of Laboratory Medicine, The First Affiliated Hospital of China Medical University, Shenyang 110001, Liaoning Province, China.
  • He X; Laboratory Medicine Innovation Unit,Chinese Academy of Medical Sciences; Department of Laboratory Medicine, The First Affiliated Hospital of China Medical University, Shenyang 110001, Liaoning Province, China.
  • Diao YY; Laboratory Medicine Innovation Unit,Chinese Academy of Medical Sciences; Department of Laboratory Medicine, The First Affiliated Hospital of China Medical University, Shenyang 110001, Liaoning Province, China,E-mail:cmudyy@163.com.
Zhongguo Shi Yan Xue Ye Xue Za Zhi ; 30(3): 851-855, 2022 Jun.
Article en Zh | MEDLINE | ID: mdl-35680816
ABSTRACT

OBJECTIVE:

Compared with the method of optical microscopy, to evaluate the accuracy of fragmented red cells(FRC) detection by Sysmex XN-3000.

METHODS:

A total of 111 samples were collected from patients diagnosed as thrombotic thrombocytopenic purpura, autoimmune disease, hematological disease, malignant tumor and health examination in our hospital from June 2019 to February 2021, including 74 cases in the case group and 37 cases in the healthy control group. All samples were detected by optical microscope and Sysmex XN-3000, respectively. ROC was used to evaluate the detection ability of Sysmex XN-3000 for schistocyte. Bland-Altman method was used to evaluate the consistency of the results of the two methods for detection of schistocyte, and Pearson correlation analysis was conducted for the difference of the results.

RESULTS:

The area under the ROC curve was 0.890(95% CI 0.828-0.952, P<0.01). Sysmex XN-3000 count did not quantitatively agree with schistocyte counts by microscopy in the case group(mean of difference-1.53, 95% limits of agreement -8.78~5.72). There was a weak positive correlation between platelet count and the difference of analyzer and microscopic results (r=0.32,P<0.05).

CONCLUSION:

Sysmex XN-3000 can be used as a reference for qualitative determination of schistocyte. However, the sensitivity of Sysmex XN-3000 should be improved. It is still necessary to combine with manual microscopy. The quantitative results are not reliable now and cannot be used as a reference for monitoring the results of schistocyte in clinical patients after treatment.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Púrpura Trombocitopénica Trombótica / Neoplasias Tipo de estudio: Diagnostic_studies / Guideline / Prognostic_studies / Qualitative_research Límite: Humans Idioma: Zh Revista: Zhongguo Shi Yan Xue Ye Xue Za Zhi Asunto de la revista: HEMATOLOGIA Año: 2022 Tipo del documento: Article País de afiliación: China

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Púrpura Trombocitopénica Trombótica / Neoplasias Tipo de estudio: Diagnostic_studies / Guideline / Prognostic_studies / Qualitative_research Límite: Humans Idioma: Zh Revista: Zhongguo Shi Yan Xue Ye Xue Za Zhi Asunto de la revista: HEMATOLOGIA Año: 2022 Tipo del documento: Article País de afiliación: China