Assessment of Blood Flow Transport in the Left Ventricle Using Ultrasound. Validation Against 4-D Flow Cardiac Magnetic Resonance.
Ultrasound Med Biol
; 48(9): 1822-1832, 2022 09.
Article
en En
| MEDLINE
| ID: mdl-35764455
ABSTRACT
Four-dimensional flow cardiac magnetic resonance (CMR) is the reference technique for analyzing blood transport in the left ventricle (LV), but similar information can be obtained from ultrasound. We aimed to validate ultrasound-derived transport in a head-to-head comparison against 4D flow CMR. In five patients and two healthy volunteers, we obtained 2D + t and 3D + t (4D) flow fields in the LV using transthoracic echocardiography and CMR, respectively. We compartmentalized intraventricular blood flow into four fractions of end-diastolic volume direct flow (DF), retained inflow (RI), delayed ejection flow (DEF) and residual volume (RV). Using ultrasound we also computed the properties of LV filling waves (percentage of LV penetration and percentage of LV volume carried by E/A waves) to determine their relationships with CMR transport. Agreement between both techniques for quantifying transport fractions was good for DF and RV (Ric [95% confidence interval] 0.82 [0.33, 0.97] and 0.85 [0.41, 0.97], respectively) and moderate for RI and DEF (Ric= 0.47 [-0.29, 0.88] and 0.55 [-0.20, 0.90], respectively). Agreement between techniques to measure kinetic energy was variable. The amount of blood carried by the E-wave correlated with DF and RV (R = 0.75 and R = 0.63, respectively). Therefore, ultrasound is a suitable method for expanding the analysis of intraventricular flow transport in the clinical setting.
Palabras clave
Texto completo:
1
Colección:
01-internacional
Base de datos:
MEDLINE
Asunto principal:
Función Ventricular Izquierda
/
Ventrículos Cardíacos
Límite:
Humans
Idioma:
En
Revista:
Ultrasound Med Biol
Año:
2022
Tipo del documento:
Article
País de afiliación:
España