Your browser doesn't support javascript.
loading
Successful implantation of a novel polymer-free everolimus-eluting stent using nitrogen-doped titanium dioxide film with good patency on follow-up angiography: A case report.
Oh, Seok; Jeong, Myung Ho; Park, Dae Sung; Kim, Munki; Kim, Jung Ha; Hyun, Dae Young; Cho, Kyung Hoon; Kim, Min Chul; Sim, Doo Sun; Hong, Young Joon; Kim, Ju Han; Ahn, Youngkeun.
Afiliación
  • Oh S; Department of Cardiology, Chonnam National University Hospital, Gwangju, Korea.
  • Jeong MH; Department of Cardiology, Chonnam National University Hospital, Gwangju, Korea.
  • Park DS; Department of Cardiology, Chonnam National University Medical School, Gwangju, Korea.
  • Kim M; Cardiovascular Research Center, Chonnam National University Hospital, Gwangju, Korea.
  • Kim JH; Cardiovascular Research Center, Chonnam National University Hospital, Gwangju, Korea.
  • Hyun DY; Cardiovascular Research Center, Chonnam National University Hospital, Gwangju, Korea.
  • Cho KH; Department of Cardiology, Chonnam National University Hospital, Gwangju, Korea.
  • Kim MC; Department of Cardiology, Chonnam National University Hospital, Gwangju, Korea.
  • Sim DS; Department of Cardiology, Chonnam National University Hospital, Gwangju, Korea.
  • Hong YJ; Department of Cardiology, Chonnam National University Medical School, Gwangju, Korea.
  • Kim JH; Department of Cardiology, Chonnam National University Hospital, Gwangju, Korea.
  • Ahn Y; Department of Cardiology, Chonnam National University Medical School, Gwangju, Korea.
Medicine (Baltimore) ; 101(29): e29666, 2022 Jul 22.
Article en En | MEDLINE | ID: mdl-35866823
RATIONALE: Despite technological advances in interventional cardiology during the last decades, many concerns remain regarding the narrowing and occlusion of the in-stent area. Particularly, polymer materials pose several problems, including chronic arterial inflammation, impaired arterial healing, and stent thrombosis. To avoid these complications, we invented the TIGEREVOLUTION stent with a cobalt-chromium alloy-based stent platform deposited with N-TiO2 film, which has demonstrated good biocompatibility. As this stent is not coated with polymer, it is expected to have decreased risk of stent thrombosis. PATIENT CONCERNS: A 62-year-old Korean man visited our department because of angina. We commenced coronary angiography (CAG). DIAGNOSIS: CAG revealed critical stenosis in the mid-portion of the right coronary artery, with a minimum lumen area of 1.08mm2 on optical coherence tomography (OCT). INTERVENTION: Percutaneous coronary intervention was performed with implantation of a novel 3.5 × 26-mm polymer-free everolimus-eluting stent using nitrogen-doped titanium dioxide film (TIGEREVOLUTION® stent). Post-percutaneous coronary intervention OCT showed good stent expansion and apposition, and the patient was discharged successfully and uneventfully. OUTCOMES: Eight months later, follow-up coronary angiography demonstrated good stent patency with no definitive evidence of in-stent restenosis, with thin stent strut coverage demonstrated on OCT. LESSONS: We report the first case of TIGEREVOLUTION stent implantation with follow-up OCT at 8 months.
Asunto(s)

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Stents Liberadores de Fármacos / Intervención Coronaria Percutánea Tipo de estudio: Observational_studies / Prognostic_studies / Risk_factors_studies Límite: Humans / Male / Middle aged Idioma: En Revista: Medicine (Baltimore) Año: 2022 Tipo del documento: Article Pais de publicación: Estados Unidos

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Stents Liberadores de Fármacos / Intervención Coronaria Percutánea Tipo de estudio: Observational_studies / Prognostic_studies / Risk_factors_studies Límite: Humans / Male / Middle aged Idioma: En Revista: Medicine (Baltimore) Año: 2022 Tipo del documento: Article Pais de publicación: Estados Unidos