Pointwise Visual Field Estimation From Optical Coherence Tomography in Glaucoma Using Deep Learning.
Hemelings, Ruben; Elen, Bart; Barbosa-Breda, João; Bellon, Erwin; Blaschko, Matthew B; De Boever, Patrick; Stalmans, Ingeborg.
Afiliación
  • Hemelings R; Research Group Ophthalmology, Department of Neurosciences, KU Leuven, Leuven, Belgium.
  • Elen B; Unit Health, Flemish Institute for Technological Research (VITO), Mol, Belgium.
  • Barbosa-Breda J; Unit Health, Flemish Institute for Technological Research (VITO), Mol, Belgium.
  • Bellon E; Research Group Ophthalmology, Department of Neurosciences, KU Leuven, Leuven, Belgium.
  • Blaschko MB; Cardiovascular R&D Center - UnIC@RISE, Department of Surgery and Physiology, Faculty of Medicine of the University of Porto, Porto, Portugal.
  • De Boever P; Department of Ophthalmology, Centro Hospitalar e Universitário São João, Porto, Portugal.
  • Stalmans I; Department of Information Technology, University Hospitals Leuven, Leuven, Belgium.
Transl Vis Sci Technol ; 11(8): 22, 2022 08 01.
Article en En | MEDLINE | ID: mdl-35998059
ABSTRACT

Purpose:

Standard automated perimetry is the gold standard to monitor visual field (VF) loss in glaucoma management, but it is prone to intrasubject variability. We trained and validated a customized deep learning (DL) regression model with Xception backbone that estimates pointwise and overall VF sensitivity from unsegmented optical coherence tomography (OCT) scans.

Methods:

DL regression models have been trained with four imaging modalities (circumpapillary OCT at 3.5 mm, 4.1 mm, and 4.7 mm diameter) and scanning laser ophthalmoscopy en face images to estimate mean deviation (MD) and 52 threshold values. This retrospective study used data from patients who underwent a complete glaucoma examination, including a reliable Humphrey Field Analyzer (HFA) 24-2 SITA Standard (SS) VF exam and a SPECTRALIS OCT.

Results:

For MD estimation, weighted prediction averaging of all four individuals yielded a mean absolute error (MAE) of 2.89 dB (2.50-3.30) on 186 test images, reducing the baseline by 54% (MAEdecr%). For 52 VF threshold values' estimation, the weighted ensemble model resulted in an MAE of 4.82 dB (4.45-5.22), representing an MAEdecr% of 38% from baseline when predicting the pointwise mean value. DL managed to explain 75% and 58% of the variance (R2) in MD and pointwise sensitivity estimation, respectively.

Conclusions:

Deep learning can estimate global and pointwise VF sensitivities that fall almost entirely within the 90% test-retest confidence intervals of the 24-2 SS test. Translational Relevance Fast and consistent VF prediction from unsegmented OCT scans could become a solution for visual function estimation in patients unable to perform reliable VF exams.
Asunto(s)

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Glaucoma / Aprendizaje Profundo Tipo de estudio: Diagnostic_studies / Observational_studies / Prognostic_studies / Risk_factors_studies Límite: Humans Idioma: En Revista: Transl Vis Sci Technol Año: 2022 Tipo del documento: Article País de afiliación: Bélgica

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Glaucoma / Aprendizaje Profundo Tipo de estudio: Diagnostic_studies / Observational_studies / Prognostic_studies / Risk_factors_studies Límite: Humans Idioma: En Revista: Transl Vis Sci Technol Año: 2022 Tipo del documento: Article País de afiliación: Bélgica