Serous body fluid evaluation using the new automated haematology analyser Mindray BC-6800Plus.
Boldú, Laura; Laguna, Javier; Casanova, Alexandra; García, Saray; Molina, Angel; Merino, Anna.
Afiliación
  • Boldú L; Haematology and Cytology Unit. CORE Laboratory. Biochemistry and Molecular Genetics Department, Biomedical Diagnostic Centre, Hospital Clinic of Barcelona-IDIBAPS, Barcelona, Spain.
  • Laguna J; Haematology and Cytology Unit. CORE Laboratory. Biochemistry and Molecular Genetics Department, Biomedical Diagnostic Centre, Hospital Clinic of Barcelona-IDIBAPS, Barcelona, Spain.
  • Casanova A; Haematology and Cytology Unit. CORE Laboratory. Biochemistry and Molecular Genetics Department, Biomedical Diagnostic Centre, Hospital Clinic of Barcelona-IDIBAPS, Barcelona, Spain.
  • García S; Haematology and Cytology Unit. CORE Laboratory. Biochemistry and Molecular Genetics Department, Biomedical Diagnostic Centre, Hospital Clinic of Barcelona-IDIBAPS, Barcelona, Spain.
  • Molina A; Haematology and Cytology Unit. CORE Laboratory. Biochemistry and Molecular Genetics Department, Biomedical Diagnostic Centre, Hospital Clinic of Barcelona-IDIBAPS, Barcelona, Spain.
  • Merino A; Haematology and Cytology Unit. CORE Laboratory. Biochemistry and Molecular Genetics Department, Biomedical Diagnostic Centre, Hospital Clinic of Barcelona-IDIBAPS, Barcelona, Spain.
Clin Chem Lab Med ; 60(11): 1786-1795, 2022 10 26.
Article en En | MEDLINE | ID: mdl-36039597
ABSTRACT

OBJECTIVES:

Cellular analysis of body fluids (BF) has clinical relevance in several medical conditions. The objective of this study is twofold (1) evaluate the analytical performance of the BF mode of Mindray BC-6800 Plus compared to manual counts under microscopy and (2) analyse if the high-fluorescent cell counts provided by the analyser (HF-BF) are useful to detect malignancy.

METHODS:

A total of 285 BF was analysed 250 corresponding to patients without neoplasia and 35 to patients with malignant diseases. Manual differential counts were performed in BF with ≥250 cells/µL. Percentages and absolute counts were obtained on the BC-6800Plus for total nucleated cells (TC-BF), mononuclear, polymorphonuclear and HF-BF. Statistical analysis was performed using Mann-Whitney U-test, Spearman's correlation, Passing-Bablok regression, Bland-Altman graph and ROC curve.

RESULTS:

To compare manual and automatic total cell counts, samples were divided in three groups <250, 250-1,000 and >1,000 cells/µL. Correlation was good in all cases (r=0.72, 0.73 and 0.92, respectively) without significant differences between both methods (p=0.65, 0.39 and 0.30, respectively). The concordance between methods showed values of 90%. Considering malignant samples, median HF-BF values showed significant higher values (102 cells/µL) with respect to non-malignant (4 cells/µL) (p<0.001). The cut-off value of 8.5 HF-BF/µL was able to discriminate samples containing malignant cells showing sensitivity and specificity values of 89 and 71%, respectively. Considering both, HF-BF and TC-BF values, sensitivity and specificity values were 100 and 53%, respectively.

CONCLUSIONS:

This study reveals that the Mindray BC-6800Plus offers an accurate and acceptable performance, showing results consistent with the manual method. It is recommended to consider both HF-BF and TC-BF values for the screening of the microscopic evaluation to ensure the detection of all malignant samples.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Líquidos Corporales / Hematología / Neoplasias Tipo de estudio: Diagnostic_studies / Guideline / Prognostic_studies Límite: Humans Idioma: En Revista: Clin Chem Lab Med Asunto de la revista: QUIMICA CLINICA / TECNICAS E PROCEDIMENTOS DE LABORATORIO Año: 2022 Tipo del documento: Article País de afiliación: España

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Líquidos Corporales / Hematología / Neoplasias Tipo de estudio: Diagnostic_studies / Guideline / Prognostic_studies Límite: Humans Idioma: En Revista: Clin Chem Lab Med Asunto de la revista: QUIMICA CLINICA / TECNICAS E PROCEDIMENTOS DE LABORATORIO Año: 2022 Tipo del documento: Article País de afiliación: España
...