Accuracy of Impression Techniques for Dental Implants Placed in Five Different Orientations.
Int J Oral Maxillofac Implants ; 37(5): 997-1002, 2022.
Article en En | MEDLINE | ID: mdl-36170315
ABSTRACT

PURPOSE:

The purpose of this study was to evaluate the impression accuracy of two implants placed in different orientations and compare the impression accuracy obtained with the dual-arch impression technique using hybrid impression copings and the conventional open-tray impression technique. MATERIALS AND

METHODS:

Five mandibular dentiform models were prepared by placing implants in the second premolar and second molar regions in the following different orientations parallel to each other; 15-degree mesiodistal angulation; 30-degree mesiodistal angulation; 15-degree buccolingual angulation; and 30-degree buccolingual angulation. After making 10 impressions for each model with the open-tray impression technique and dual-arch impression technique with hybrid impression copings, cast models were fabricated for each impression (n = 10). Scan bodies were mounted on the five dentiform models and the fabricated 100 cast models were scanned using a desktop scanner. The three-dimensional deviation of the scan bodies on the cast models was calculated and compared to the reference data from the dentiform models. A two-way analysis of variance was conducted (α = .05).

RESULTS:

The root mean square deviation values obtained from the implants placed with 30-degree mesiodistal angulation were 93.05 ± 6.21 µm with the open-tray impression technique and 104.01 ± 8.89 µm with the dual-arch impression technique, which were the largest deviation values for both techniques (P < .001). Compared to the open-tray impression technique, the dual-arch impression technique with the hybrid impression copings showed significantly lower accuracy when the angulation between the implants was 15 degrees mesiodistally (P < .001), 30 degrees mesiodistally (P = .016), or 30 degrees buccolingually (P < .001). However, there was no significant difference between the accuracy of the two impression techniques for parallel implants (P = .74).

CONCLUSION:

When the two implants were inclined 30 degrees mesiodistally, both implant impression techniques showed the largest deviation and the dual-arch impression technique showed lower accuracy compared to the conventional open-tray impression technique. Parallel placement of implants may improve impression accuracy and enable use of the dual-arch impression technique.