Evaluation of the interchangeability between the new fully-automated affinity chrome-mediated immunoassay (ACMIA) and the Quantitative Microsphere System (QMS) with a CE-IVD-certified LC-MS/MS assay for therapeutic drug monitoring of everolimus after solid organ transplantation.
Ialongo, Cristiano; D'alessandro, Annamaria; Sapio, Maria; Angeloni, Antonio; Porzio, Ottavia.
Afiliación
  • Ialongo C; Department of Experimental Medicine, Policlinico Umberto I, "Sapienza" University, Rome, Italy.
  • D'alessandro A; Clinical Biochemistry Laboratory, IRCCS "Bambino Gesù" Children's Hospital, Rome, Italy.
  • Sapio M; Department of Experimental Medicine, Policlinico Umberto I, "Sapienza" University, Rome, Italy.
  • Angeloni A; Department of Experimental Medicine, Policlinico Umberto I, "Sapienza" University, Rome, Italy.
  • Porzio O; Clinical Biochemistry Laboratory, IRCCS "Bambino Gesù" Children's Hospital, Rome, Italy.
Clin Chem Lab Med ; 61(2): 245-250, 2023 01 27.
Article en En | MEDLINE | ID: mdl-36330751
ABSTRACT

OBJECTIVES:

This study aims to evaluate the interchangeability between the Siemens Healthineers' "EVRO" new affinity chrome-mediated immunoassay (ACMIA/EVRO) and Thermo Fisher Scientific's "EVER" Quantitative Microsphere System (QMS/EVER) with Chromsystems' CE-IVD-certified "MassTox" liquid-chromatography/tandem-mass spectrometry (LC-MS/MS) assay for the therapeutic drug monitoring of everolimus.

METHODS:

A single lot of reagent, calibrators and controls were used for each assay. A total of 67 whole blood samples (n=67) from patients receiving solid organ transplant were analyzed (n=31 with kidney transplant and n=36 with liver transplant); Passing-Bablok regression and Bland-Altman difference plot were used to evaluate bias and individual agreement; LC-MS/MS analysis was used to measure the actual concentrations of calibrators and controls compared to the assigned value.

RESULTS:

ACMIA/EVRO did not show any systematic bias compared to LC-MS/MS (intercept=0.244 ng/mL, 95% CI -0.254 to 0.651 ng/mL). Nevertheless, significant proportional bias (slope=1.511, 95% CI 1.420 to 1.619) associated to a combined bias of 44.8% (95% CI 41.2-48.3%) was observed. Conversely, QMS/EVER did not show any bias at both systematic (intercept=-0.151 ng/mL, 95% CI -0.671 to 0.256 ng/mL) and proportional level (slope=0.971, 95% CI 0.895 to 1.074) with a non-statistically significant combined bias of -3.6% (95% CI -8.4-1.1%). Based on a concentration of calibrators and controls above the assigned value for both the analytical methods, in the ACMIA/EVRO a correction which was approximately one-third of the correction for the QMS/EVER was observed.

CONCLUSIONS:

ACMIA/EVRO but not QMS/EVER shows a lack of interchangeability with the CE-IVD-certified LC-MS/MS assay. We hypothesize that, as the ACMIA/EVRO uses an anti-sirolimus antibody, the under-corrected assigned value in the assay calibrators was not sufficient to reproduce the everolimus metabolites cross-reactivity occurring in real samples.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Trasplante de Órganos / Everolimus Límite: Humans Idioma: En Revista: Clin Chem Lab Med Asunto de la revista: QUIMICA CLINICA / TECNICAS E PROCEDIMENTOS DE LABORATORIO Año: 2023 Tipo del documento: Article País de afiliación: Italia

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Trasplante de Órganos / Everolimus Límite: Humans Idioma: En Revista: Clin Chem Lab Med Asunto de la revista: QUIMICA CLINICA / TECNICAS E PROCEDIMENTOS DE LABORATORIO Año: 2023 Tipo del documento: Article País de afiliación: Italia