Your browser doesn't support javascript.
loading
Does the fixture thread depth affect the accuracy of implant placement during fully guided immediate implant placement?: A human cadaver study.
Song, Young Woo; Park, Jin-Young; Jung, Ji-Young; Kim, Joo-Nyeon; Hu, Kyung-Seok; Lee, Jung-Seok.
Afiliación
  • Song YW; Department of Periodontology, Research Institute for Periodontal Regeneration, Yonsei University College of Dentistry, Seoul, South Korea.
  • Park JY; Innovation Research and Support Center for Dental Science, Yonsei University Dental Hospital, Seoul, South Korea.
  • Jung JY; Department of Periodontology, Research Institute for Periodontal Regeneration, Yonsei University College of Dentistry, Seoul, South Korea.
  • Kim JN; Innovation Research and Support Center for Dental Science, Yonsei University Dental Hospital, Seoul, South Korea.
  • Hu KS; Department of Periodontology, Research Institute for Periodontal Regeneration, Yonsei University College of Dentistry, Seoul, South Korea.
  • Lee JS; Innovation Research and Support Center for Dental Science, Yonsei University Dental Hospital, Seoul, South Korea.
Clin Oral Implants Res ; 34(2): 116-126, 2023 Feb.
Article en En | MEDLINE | ID: mdl-36458928
ABSTRACT

BACKGROUND:

The fixture thread depth reportedly influences the primary stability of dental implant, but its effect on the positional accuracy in immediate implant placement has not been reported previously. MATERIALS AND

METHODS:

Fifty-six single-rooted, anterior and premolar teeth were extracted from six human cadavers, followed by installing either regular-threaded implants (RT group, N = 30) or deep-threaded implants (DT group, N = 26) completely relying on the surgical guide. Optical impressions taken after osteotomy and fixture installation were superimposed with the preoperative virtual planning data to measure the vertical, angular, platform and apex deviations that occurred during osteotomy and installation.

RESULTS:

While the osteotomy deviations were similar between the two groups, the angular and apex deviation of the DT group (2.67 ± 2.56°; 1.04 ± 0.49 mm, respectively) were significantly larger than those of the RT group (1.61 ± 1.04°; 0.67 ± 0.41 mm, respectively) during installation (p < .05). When the installation deviations were analysed in the anterior and premolar areas separately, the angular, platform and apex deviations of the DT group (3.05 ± 3.26°; 0.75 ± 0.32 mm; 1.08 ± 0.56 mm, respectively) were significantly larger than those of the RT group (1.56 ± 0.73°, 0.59 ± 0.28 mm; 0.62 ± 0.38 mm, respectively) in the anterior site (p < .05), whereas there was no significant intergroup difference in the premolar site.

CONCLUSION:

In immediate-implant surgery, using an implant with a deeper thread might increase the deviations during installation, especially in the anterior area.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Implantes Dentales / Cirugía Asistida por Computador Límite: Humans Idioma: En Revista: Clin Oral Implants Res Asunto de la revista: ODONTOLOGIA Año: 2023 Tipo del documento: Article País de afiliación: Corea del Sur

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Implantes Dentales / Cirugía Asistida por Computador Límite: Humans Idioma: En Revista: Clin Oral Implants Res Asunto de la revista: ODONTOLOGIA Año: 2023 Tipo del documento: Article País de afiliación: Corea del Sur