Brain imaging of sequential acquisition using a flexible PET scanner and 3-T MRI: quantitative and qualitative assessment.
Ann Nucl Med
; 37(4): 209-218, 2023 Apr.
Article
en En
| MEDLINE
| ID: mdl-36585566
OBJECTIVE: A mobile PET scanner termed flexible PET (fxPET) has been designed to fit existing MRI systems. The purpose of this study was to assess brain imaging with fxPET combined with 3-T MRI in comparison with conventional PET (cPET)/CT. METHODS: In this prospective study, 29 subjects with no visible lesions except for mild leukoaraiosis on whole brain imaging underwent 2-deoxy-2-[18F]fluoro-D-glucose ([18F]FDG) cPET/CT followed by fxPET and MRI. The registration differences between fxPET and MRI and between cPET and CT were compared by measuring spatial coordinates. Three-dimensional magnetization-prepared rapid acquisition gradient-echo T1-weighted imaging (T1WI) was acquired. We applied two methods of attenuation correction to the fxPET images: MR-based attenuation correction, which yielded fxPETMRAC; and CT-based attenuation correction, which yielded fxPETCTAC. The three PET datasets were co-registered to the T1WI. Following subcortical segmentation and cortical parcellation, volumes of interest were placed in each PET image to assess physiological accumulation in the brain. SUVmean was obtained and compared between the three datasets. We also visually evaluated image distortion and clarity of fxPETMRAC. RESULTS: Mean misregistration of fxPET/MRI was < 3 mm for each margin. Mean registration differences were significantly larger for fxPET/MRI than for cPET/CT except for the superior margin. There were high correlations between the three PET datasets regarding SUVmean. On visual evaluation of image quality, the grade of distortion was comparable between fxPETMRAC and cPET, and the grade of clarity was acceptable but inferior for fxPETMRAC compared with cPET. CONCLUSIONS: fxPET could successfully determine physiological [18F]FDG uptake; however, improved image clarity is desirable. In this study, fxPET/MRI at 3-T was feasible for brain imaging.
Palabras clave
Texto completo:
1
Colección:
01-internacional
Base de datos:
MEDLINE
Asunto principal:
Fluorodesoxiglucosa F18
/
Imagen Multimodal
Tipo de estudio:
Observational_studies
/
Qualitative_research
Límite:
Humans
Idioma:
En
Revista:
Ann Nucl Med
Asunto de la revista:
MEDICINA NUCLEAR
Año:
2023
Tipo del documento:
Article
País de afiliación:
Japón
Pais de publicación:
Japón