Your browser doesn't support javascript.
loading
Dose Agreement Analysis of Treatment Planning System-Calculated Doses and Markus Chamber-Measured Doses in the Near-Surface Region for Breast Cancer Patients' Conformal Treatment Plans.
Gaur, Garima; Banipal, Raja Paramjeet Singh; Garg, Pardeep; Gurjar, Om Prakash; Kaur, Gurpreet; Sachdeva, Sheetal; Grover, Romikant; Kang, Manraj Singh; Dangwal, Vinod Kumar.
Afiliación
  • Gaur G; Department of Radiation Oncology, Guru Gobind Singh Medical College and Hospital, Faridkot, Punjab, India.
  • Banipal RPS; Department of Radiation Oncology, Government Medical College, Patiala, Punjab, India.
  • Garg P; Department of Radiation Oncology, Guru Gobind Singh Medical College and Hospital, Faridkot, Punjab, India.
  • Gurjar OP; Department of Radiotherapy, Government Cancer Hospital, MGM Medical College, Indore, Madhya Pradesh, India.
  • Kaur G; Department of Radiation Oncology, Guru Gobind Singh Medical College and Hospital, Faridkot, Punjab, India.
  • Sachdeva S; Advance Cancer Centre, Bhatinda, Punjab, India.
  • Grover R; Department of Radiation Oncology, Guru Gobind Singh Medical College and Hospital, Faridkot, Punjab, India.
  • Kang MS; Department of Radiation Oncology, Guru Gobind Singh Medical College and Hospital, Faridkot, Punjab, India.
  • Dangwal VK; Department of Radiation Oncology, Government Medical College, Patiala, Punjab, India.
J Med Phys ; 47(3): 256-261, 2022.
Article en En | MEDLINE | ID: mdl-36684697
ABSTRACT

Background:

Surface/skin dose measurement is one of the most challenging tasks for clinical dosimetry in radiotherapy and comparison with almost all the commercially available treatment planning systems (TPSs) brings a significant variation with the measured dose. Aims and

Objectives:

In the current study, doses calculated from the TPS in the near-surface region for conformal plans (both three-dimensional conformal radiotherapy [3DCRT] and intensity-modulated radiotherapy [IMRT]) of 35 breast cancer patients were evaluated and compared with the doses measured with Markus chamber. Materials and

Methods:

The computed tomography (CT) images of a solid water slab phantom with a Markus chamber (at different depths ranging from 1 mm to 5 mm from the surface) were taken and imported into the TPS. All the conformal treatment plans made in TPS were executed on a linear accelerator and dose agreements between TPS calculated and chamber measured doses were analysed.

Results:

Results showed that this TPS underestimated the calculated doses in the superficial region by up to 26% and 21%, respectively, with respect to mean and maximum dose values obtained within the effective volume of the chamber used.

Conclusion:

The uncertainty of doses in the superficial region should be kept in mind when evaluating treatment plans for superficial tumours in TPS.
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Idioma: En Revista: J Med Phys Año: 2022 Tipo del documento: Article País de afiliación: India

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Idioma: En Revista: J Med Phys Año: 2022 Tipo del documento: Article País de afiliación: India