Accuracy of injection pressure measurement at peripheral nerves using high-resolution 40 MHz ultrasound in an anesthetized porcine model.
McLeod, Graeme A; Cowie, Alistair; Sadler, Amy; Watson, Fiona; Wasik, Paul; Reina, Miguel Angel.
Afiliación
  • McLeod GA; Department of Anaesthesia, Ninewells Hospital, Dundee, UK g.a.mcleod@dundee.ac.uk.
  • Cowie A; Institute of Academic Anaesthesia, University of Dundee, Dundee, UK.
  • Sadler A; University of Dundee School of Medicine, Dundee, UK.
  • Watson F; Department of Anaesthesia, NHS Tayside, Dundee, UK.
  • Wasik P; Department of Anaesthesia, NHS Tayside, Dundee, UK.
  • Reina MA; Department of Anaesthesia, NHS Tayside, Dundee, UK.
Reg Anesth Pain Med ; 48(10): 501-507, 2023 10.
Article en En | MEDLINE | ID: mdl-36822816
ABSTRACT

BACKGROUND:

Fluid injection pressure measurement is promoted as a marker of needle tip position that discriminates between tissue layers. However, clinical ultrasound has insufficient resolution to identify the exact position of the needle tip. Our primary objective was to use 40 MHz ultrasound in anesthetized pigs in order to precisely locate the tip of the needle and measure opening injection pressure in muscle, at epineurium and in subepineurium.

METHODS:

We surgically exposed the axillae of four anesthetized pigs. Two operators placed a 40 MHz ultrasound transducer over the pectoral muscle and imaged axillary, median and radial nerves. Injections (0.5 mL) were randomized to in-plane and out-of-plane needle trajectories and flow rates of 1, 6 and 12 mL/min.

RESULTS:

We identified 541 fascicles in 23 nerves. The ratio of fascicle area to nerve area remained constant at ~0.30 for all nerves. Axillary nerves were smaller than median and radial nerves, difference in diameter (95% CI) 1.61 (0.87 to 2.36) mm, p<0.001 and 1.59 (0.82 to 2.36) mm, p=0.001, respectively. Axillary nerves had less fascicles per nerve than median nerves, difference 7.63 (2.43 to 12.83) and radial nerves, difference 9.02 (3.64 to 14.40). We visualized the circumneurium and injection within the subcircumneural compartment. Intraneural injection increased nerve area (SD) from 5.7 (2.2) mm2 to 13.7 (5.5) mm2, difference 8.0 (5.4-10.6) mm2, p<0.001. Mean injection pressure was greater in subepineurium compared with muscle, geometric ratio 2.29 (1.30 to 4.10), p<0.001; and greater on epineurium compared with muscle, geometric ratio 1.73 (1.03 to 3.00), p=0.01. Twenty-two out of 23 injections in muscle, 14 out of 23 injections at epineurium and 11 out of 22 injections in subepineurium were <138 kPa (20 psi).

CONCLUSION:

Needle tip position was not discernible using pressure monitoring. The circumneurium and subcircumneural injection compartment were observed but not intrafascicular injection.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Bloqueo Nervioso Tipo de estudio: Clinical_trials / Diagnostic_studies Límite: Animals / Humans Idioma: En Revista: Reg Anesth Pain Med Asunto de la revista: ANESTESIOLOGIA / NEUROLOGIA / PSICOFISIOLOGIA Año: 2023 Tipo del documento: Article País de afiliación: Reino Unido

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Bloqueo Nervioso Tipo de estudio: Clinical_trials / Diagnostic_studies Límite: Animals / Humans Idioma: En Revista: Reg Anesth Pain Med Asunto de la revista: ANESTESIOLOGIA / NEUROLOGIA / PSICOFISIOLOGIA Año: 2023 Tipo del documento: Article País de afiliación: Reino Unido