Imprecision remains to be improved in the measurement of serum cystatin C with heterogeneous systems.
Zeng, Jie; Zhang, Li; Zhang, Jiangtao; Zhou, Weiyan; Zhang, Tianjiao; Wang, Jing; Zhao, Haijian; Zhang, Chuanbao.
Afiliación
  • Zeng J; National Center for Clinical Laboratories, Chinese Academy of Medical Sciences, Beijing Hospital/National Center of Gerontology, Beijing Engineering Research Center of Laboratory Medicine, Institute of Geriatric Medicine, Beijing, P.R. China.
  • Zhang L; National Center for Clinical Laboratories, Chinese Academy of Medical Sciences, Beijing Hospital/National Center of Gerontology, Beijing Engineering Research Center of Laboratory Medicine, Chinese Academy of Medical Sciences and Peking Union Medical College, Institute of Geriatric Medicine, Beijing,
  • Zhang J; National Center for Clinical Laboratories, Chinese Academy of Medical Sciences, Beijing Hospital/National Center of Gerontology, Beijing Engineering Research Center of Laboratory Medicine, Institute of Geriatric Medicine, Beijing, P.R. China.
  • Zhou W; National Center for Clinical Laboratories, Chinese Academy of Medical Sciences, Beijing Hospital/National Center of Gerontology, Beijing Engineering Research Center of Laboratory Medicine, Institute of Geriatric Medicine, Beijing, P.R. China.
  • Zhang T; National Center for Clinical Laboratories, Chinese Academy of Medical Sciences, Beijing Hospital/National Center of Gerontology, Beijing Engineering Research Center of Laboratory Medicine, Institute of Geriatric Medicine, Beijing, P.R. China.
  • Wang J; National Center for Clinical Laboratories, Chinese Academy of Medical Sciences, Beijing Hospital/National Center of Gerontology, Beijing Engineering Research Center of Laboratory Medicine, Institute of Geriatric Medicine, Beijing, P.R. China.
  • Zhao H; National Center for Clinical Laboratories, Chinese Academy of Medical Sciences, Beijing Hospital/National Center of Gerontology, Beijing Engineering Research Center of Laboratory Medicine, Institute of Geriatric Medicine, Beijing, P.R. China.
  • Zhang C; National Center for Clinical Laboratories, Chinese Academy of Medical Sciences, Beijing Hospital/National Center of Gerontology, Beijing Engineering Research Center of Laboratory Medicine, Institute of Geriatric Medicine, Beijing, P.R. China.
Clin Chem Lab Med ; 61(8): 1455-1462, 2023 07 26.
Article en En | MEDLINE | ID: mdl-36866730
ABSTRACT

OBJECTIVES:

Except for the large bias of some measurement systems for serum cystatin C (CysC) measurements, unacceptable imprecision has been observed for the heterogenous system. This study analyzed the external quality assessment (EQA) results in 2018-2021 to provide an insight into the imprecision of CysC assays.

METHODS:

Five EQA samples were sent to participating laboratories every year. Participants were divided into reagent/calibrator-based peer groups, for which the robust mean of each sample and robust coefficient of variation (CV) were calculated by Algorithm A from ISO 13528. Peers with more than 12 participants per year were selected for further analysis. The limit of CV was determined to be 4.85% based on clinical application requirements. The concentration-related effect on CVs was investigated using logarithmic curve fitting; the difference in medians and robust CVs between instrument-based subgroups was also evaluated.

RESULTS:

The total number of participating laboratories increased from 845 to 1,695 in four years and heterogeneous systems remained the mainstream (≥85%). Of 18 peers with ≥12 participants, those using homogeneous systems showed relatively steady and small CVs over four years, with the mean four-year CVs ranging from 3.21 to 3.68%. Some peers using heterogenous systems showed reduced CVs over four years, while 7/15 still had unacceptable CVs in 2021 (5.01-8.34%). Six peers showed larger CVs at the low or high concentrations, and some instrument-based subgroups presented greater imprecision than others.

CONCLUSIONS:

More efforts should be made to improve the imprecision of heterogeneous systems for CysC measurement.