Outcomes After Revision Total Knee Arthroplasty From a Specific, Now-Recalled Implant System.
Puri, Simarjeet; Chandi, Sonia; Chiu, Yu-Fen; Blevins, Jason L; Westrich, Geoffrey H; Figgie, Mark P; Sculco, Peter K; Chalmers, Brian P; Gausden, Elizabeth B.
Afiliación
  • Puri S; Department of Orthopedic Surgery, Hospital for Special Surgery, New York, New York.
  • Chandi S; Department of Orthopedic Surgery, Hospital for Special Surgery, New York, New York.
  • Chiu YF; Biostatistics Core, Hospital for Special Surgery, New York, New York.
  • Blevins JL; Department of Orthopedic Surgery, Hospital for Special Surgery, New York, New York.
  • Westrich GH; Department of Orthopedic Surgery, Hospital for Special Surgery, New York, New York.
  • Figgie MP; Department of Orthopedic Surgery, Hospital for Special Surgery, New York, New York.
  • Sculco PK; Department of Orthopedic Surgery, Hospital for Special Surgery, New York, New York.
  • Chalmers BP; Department of Orthopedic Surgery, Hospital for Special Surgery, New York, New York.
  • Gausden EB; Department of Orthopedic Surgery, Hospital for Special Surgery, New York, New York.
J Arthroplasty ; 38(6S): S290-S296.e1, 2023 06.
Article en En | MEDLINE | ID: mdl-36907386
ABSTRACT

BACKGROUND:

A specific total knee arthroplasty (TKA) implant system was recently recalled for high incidence of early polymeric wear and osteolysis. We analyzed the early outcomes of aseptic revision with these implants.

METHODS:

We identified 202 aseptic revision TKAs of this implant system performed at a single institution between 2010 and 2020. Revision indications included aseptic loosening (n = 120), instability (n = 55), and polymeric wear/osteolysis (n = 27). Components were revised in 145 cases (72%) and isolated polyethylene insert exchange occurred in 57 cases (28%). Kaplan-Meier and Cox proportional hazards analyses were used to determine survivorship free from all-cause rerevision and rerevision risk factors.

RESULTS:

At 2 and 5 years, survivorship free from all-cause rerevision was 89 and 76% in the polyethylene exchange cohort versus 92 and 84% in the component revision cohort (P = .5). At 2 and 5 years, survivorship in revision with components from the same manufacturer was 89 and 80% compared to 95 and 86% in revision with components from a different manufacturer (P = .2). Among re-revisions (n = 30), cones (37%), sleeves (7%), hinge/distal femoral replacement implants were frequently used (13%). Men had increased risk for rerevision (hazard ratio = 2.3, P = .04).

CONCLUSION:

In this series of aseptic revision TKAs performed on a now-recalled implant system, survivorship free from rerevision was lower than expected when components from the same manufacturer were utilized, but comparable to contemporary reports when both the components were revised with an alternative implant system. Metaphyseal fixation with cones and sleeves as well as highly constrained implants was frequently utilized at time of rerevision TKA. LEVEL OF EVIDENCE Level IV.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Osteólisis / Artroplastia de Reemplazo de Rodilla / Prótesis de la Rodilla Tipo de estudio: Etiology_studies / Observational_studies / Prognostic_studies / Risk_factors_studies Límite: Humans / Male Idioma: En Revista: J Arthroplasty Asunto de la revista: ORTOPEDIA Año: 2023 Tipo del documento: Article

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Osteólisis / Artroplastia de Reemplazo de Rodilla / Prótesis de la Rodilla Tipo de estudio: Etiology_studies / Observational_studies / Prognostic_studies / Risk_factors_studies Límite: Humans / Male Idioma: En Revista: J Arthroplasty Asunto de la revista: ORTOPEDIA Año: 2023 Tipo del documento: Article
...