Your browser doesn't support javascript.
loading
Gestational age assessed by optical skin reflection in low-birth-weight newborns: Applications in classification at birth.
Vitral, Gabriela Luiza Nogueira; Romanelli, Roberta Maia de Castro; Reis, Zilma Silveira Nogueira; Guimarães, Rodney Nascimento; Dias, Ivana; Mussagy, Nilza; Taunde, Sergio; Neves, Gabriela Silveira; de São José, Carolina Nogueira; Pantaleão, Alexandre Negrão; Pappa, Gisele Lobo; Gaspar, Juliano de Souza; de Aguiar, Regina Amélia Pessoa Lopes.
Afiliación
  • Vitral GLN; Faculty of Medicine, Universidade Federal de Minas Gerais, Belo Horizonte, Brazil.
  • Romanelli RMC; Faculdade Ciências Médicas de Minas Gerais, Belo Horizonte, Brazil.
  • Reis ZSN; Faculty of Medicine, Universidade Federal de Minas Gerais, Belo Horizonte, Brazil.
  • Guimarães RN; Faculty of Medicine, Universidade Federal de Minas Gerais, Belo Horizonte, Brazil.
  • Dias I; Faculty of Medicine, Universidade Federal de Minas Gerais, Belo Horizonte, Brazil.
  • Mussagy N; Hospital Central de Maputo, Maputo, Mozabique.
  • Taunde S; Hospital Central de Maputo, Maputo, Mozabique.
  • Neves GS; Hospital Central de Maputo, Maputo, Mozabique.
  • de São José CN; Faculty of Medicine, Universidade Federal de Minas Gerais, Belo Horizonte, Brazil.
  • Pantaleão AN; Hospital Sofia Feldman, Belo Horizonte, Brazil.
  • Pappa GL; Faculdade Ciências Médicas de Minas Gerais, Belo Horizonte, Brazil.
  • Gaspar JS; Faculty of Medicine, Universidade Federal de Minas Gerais, Belo Horizonte, Brazil.
  • de Aguiar RAPL; Computer Science Department, Universidade Federal de Minas Gerais, Belo Horizonte, Brazil.
Front Pediatr ; 11: 1141894, 2023.
Article en En | MEDLINE | ID: mdl-37056944
ABSTRACT

Introduction:

A new medical device was previously developed to estimate gestational age (GA) at birth by processing a machine learning algorithm on the light scatter signal acquired on the newborn's skin. The study aims to validate GA calculated by the new device (test), comparing the result with the best available GA in newborns with low birth weight (LBW).

Methods:

We conducted a multicenter, non-randomized, and single-blinded clinical trial in three urban referral centers for perinatal care in Brazil and Mozambique. LBW newborns with a GA over 24 weeks and weighing between 500 and 2,500 g were recruited in the first 24 h of life. All pregnancies had a GA calculated by obstetric ultrasound before 24 weeks or by reliable last menstrual period (LMP). The primary endpoint was the agreement between the GA calculated by the new device (test) and the best available clinical GA, with 95% confidence limits. In addition, we assessed the accuracy of using the test in the classification of preterm and SGA. Prematurity was childbirth before 37 gestational weeks. The growth standard curve was Intergrowth-21st, with the 10th percentile being the limit for classifying SGA.

Results:

Among 305 evaluated newborns, 234 (76.7%) were premature, and 139 (45.6%) were SGA. The intraclass correlation coefficient between GA by the test and reference GA was 0.829 (95% CI 0.785-0.863). However, the new device (test) underestimated the reference GA by an average of 2.8 days (95% limits of agreement -40.6 to 31.2 days). Its use in classifying preterm or term newborns revealed an accuracy of 78.4% (95% CI 73.3-81.6), with high sensitivity (96.2%; 95% CI 92.8-98.2). The accuracy of classifying SGA newborns using GA calculated by the test was 62.3% (95% CI 56.6-67.8).

Discussion:

The new device (test) was able to assess GA at birth in LBW newborns, with a high agreement with the best available GA as a reference. The GA estimated by the device (test), when used to classify newborns on the first day of life, was useful in identifying premature infants but not when applied to identify SGA infants, considering current algohrithm. Nonetheless, the new device (test) has the potential to provide important information in places where the GA is unknown or inaccurate.
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Tipo de estudio: Clinical_trials Idioma: En Revista: Front Pediatr Año: 2023 Tipo del documento: Article País de afiliación: Brasil

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Tipo de estudio: Clinical_trials Idioma: En Revista: Front Pediatr Año: 2023 Tipo del documento: Article País de afiliación: Brasil