Your browser doesn't support javascript.
loading
Estimating size specific dose estimate from computed tomography radiograph localizer with radiation risk assessment.
Burton, Christiane Sarah; Al-Ward, Shahad.
Afiliación
  • Burton CS; Department of Diagnostic Imaging, St Jude Children's Research Hospital, Memphis, Tennessee, USA.
  • Al-Ward S; Department of Radiation Oncology, St Jude Children's Research Hospital, Memphis, Tennessee, USA.
J Appl Clin Med Phys ; 24(6): e13989, 2023 Jun.
Article en En | MEDLINE | ID: mdl-37132289
ABSTRACT

BACKGROUND:

Quantifying radiation burden is necessary for optimizing imaging protocols. The normalized dose coefficient (NDC) is determined from the water-equivalent diameter (WED) and is used to scale the CTDIvol based on body habitus to determine the size specific dose estimate (SSDE). In this study we determine the SSDE prior to the CT scan and how sensitive the SSDE from WED is to the lifetime attributable risk (LAR) from BEIR VII.

METHOD:

For calibration, phantom images are used to relate the mean pixel values along a profile ( PPV ¯ $\overline {{\rm{PPV}}} $ ) of the CT localizer to the water-equivalent area (AW ) of the CT axial scan at the same z-location. Images of the CTDIvol phantoms (32 cm, 16 cm, and ∼1 cm) and ACR phantom (Gammex 464) were acquired on four scanners. The relationship between the AW and PPV ¯ $\overline {{\rm{PPV}}} $ was used to calculate the WED from the CT localizer for patient scans. A total of 790 CT examinations of the chest and abdominopelvic regions were used in this study. The effective diameter (ED) was calculated from the CT localizer. The LAR was calculated based on the patient chest and abdomen using the National Cancer Institute Dosimetry System for Computed Tomography (NCICT). The radiation sensitivity index (RSI) and risk differentiability index (RDI) were calculated for SSDE and CTDIvol.

RESULTS:

The WED from CT localizers and CT axials scans show good correlation (R2  = 0.96) with the maximum percentage difference being 13.45%. The NDC from WED correlates poorly with LAR for lungs (R2  = 0.18) and stomach (R2  = 0.19), however that is the best correlation.

CONCLUSION:

The SSDE can be determined within 20% as recommended by the report of AAPM TG 220. The CTDIvol and SSDE are not good surrogates for radiation risk, however the sensitivity for SSDE improves when using WED instead of ED.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Radiometría / Tomografía Computarizada por Rayos X Tipo de estudio: Etiology_studies / Risk_factors_studies Límite: Humans Idioma: En Revista: J Appl Clin Med Phys Asunto de la revista: BIOFISICA Año: 2023 Tipo del documento: Article País de afiliación: Estados Unidos

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Radiometría / Tomografía Computarizada por Rayos X Tipo de estudio: Etiology_studies / Risk_factors_studies Límite: Humans Idioma: En Revista: J Appl Clin Med Phys Asunto de la revista: BIOFISICA Año: 2023 Tipo del documento: Article País de afiliación: Estados Unidos