Mechanical deformation of elastomer medical devices can enable microbial surface colonization.
van den Berg, Desmond; Asker, Dalal; Awad, Tarek S; Lavielle, Nicolas; Hatton, Benjamin D.
Afiliación
  • van den Berg D; Institute of Biomedical Engineering, University of Toronto, Toronto, Canada.
  • Asker D; Department of Materials Science and Engineering, University of Toronto, Toronto, Canada.
  • Awad TS; Food Science & Technology Department, Alexandria University, Alexandria, Egypt.
  • Lavielle N; Department of Materials Science and Engineering, University of Toronto, Toronto, Canada.
  • Hatton BD; Physique et Mécanique des Milieux Hétérogènes, CNRS, ESPCI, PSL Research University, Sorbonne Université, Sorbonne Paris Cité, 75005, Paris, France.
Sci Rep ; 13(1): 7691, 2023 05 11.
Article en En | MEDLINE | ID: mdl-37169828
ABSTRACT
Elastomers such as silicone are common in medical devices (catheters, prosthetic implants, endoscopes), but they remain prone to microbial colonization and biofilm infections. For the first time, our work shows that rates of microbial surface attachment to polydimethylsiloxane (PDMS) silicone can be significantly affected by mechanical deformation. For a section of bent commercial catheter tubing, bacteria (P. aeruginosa) show a strong preference for the 'convex' side compared to the 'concave' side, by a factor of 4.2. Further testing of cast PDMS materials in bending only showed a significant difference for samples that were manually wiped (damaged) beforehand (1.75 × 104 and 6.02 × 103 cells/mm2 on the convex and concave sides, respectively). We demonstrate that surface microcracks in elastomers are opened under tensile stress (convex bending) to become 'activated' as sites for microbial colonization. This work demonstrates that the high elastic limit of elastomers enables these microcracks to reversibly open and close, as 'dynamic defects'. Commercial catheters have relatively high surface roughness inherent to manufacturing, but we show that even manual wiping of newly-cast PDMS is sufficient to generate surface microcracks. We consider the implication for medical devices that feature sustained, surgical, or cyclic deformation, in which localized tensile conditions may expose these surface defects to opportunistic microbes. As a result, our work showcases serious potential problems in the widespread usage and development of elastomers in medical devices.