Sensitive Measurement of Clinically Relevant Factor VIII Levels in Thrombin Generation Assays Requires Presence of Factor XIa.
van de Berg, Tom W; Beckers, Erik A M; Heubel-Moenen, Floor C J I; Henskens, Yvonne M C; Thomassen, M Christella L G D; Hackeng, Tilman M.
Afiliación
  • van de Berg TW; Department of Biochemistry, Cardiovascular Research Institute Maastricht (CARIM), University Maastricht, Maastricht, The Netherlands.
  • Beckers EAM; Department of Hematology, Division of Internal Medicine, Maastricht UMC + , Maastricht, The Netherlands.
  • Heubel-Moenen FCJI; Department of Hematology, Division of Internal Medicine, Maastricht UMC + , Maastricht, The Netherlands.
  • Henskens YMC; Department of Hematology, Division of Internal Medicine, Maastricht UMC + , Maastricht, The Netherlands.
  • Thomassen MCLGD; Central Diagnostics Laboratory, Maastricht UMC + , Maastricht, The Netherlands.
  • Hackeng TM; Department of Biochemistry, Cardiovascular Research Institute Maastricht (CARIM), University Maastricht, Maastricht, The Netherlands.
Thromb Haemost ; 123(11): 1034-1041, 2023 Nov.
Article en En | MEDLINE | ID: mdl-37236229
ABSTRACT

BACKGROUND:

Hemophilia A (HA) is characterized by decreased or absent factor VIII (FVIII) activity. Current FVIII assays are based on clotting time and thus only provide information about the initiation of coagulation. In contrast, thrombin generation assays (TGAs) can be used to measure the full coagulation spectrum of initiation, propagation, and termination that provide information on the whole course of thrombin generation and inhibition. However, the commercially available TG kits lack sensitivity for measurements of hemophilia plasma within lower FVIII ranges, which is essential for explaining differences in bleeding phenotypes in hemophiliacs at clinically low levels of FVIII.

AIMS:

Optimization of the TGA for measurements of low FVIII levels in severe HA patients.

METHODS:

TGA measurements were performed in severe HA pooled plasma (n = 10). Investigations of several preanalytical and analytical variables of the assay were performed in a stepwise process and adjusted based on sensitivity toward intrinsic coagulation activation.

RESULTS:

TGA initiated by tissue factor (TF) alone at varying concentrations was unable to significantly differentiate between FVIII levels below 20%. In contrast, TGA activation with low concentrations of TF in presence of FXIa appeared to be highly sensitive for FVIII changes both in high and low ranges. In addition, a representative TGA curve at trough levels could only be produced using the dual TF/FXIa TGA.

CONCLUSION:

We propose a critical optimization for the setup of the TGA for measurements in severe HA plasma. The dual TF/FXIa TGA shows increased sensitivity, especially in lower FVIII ranges, which allows for better individual characterization at baseline, prediction of interventions, and follow-up.
Asunto(s)

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Hemostáticos / Hemofilia A Tipo de estudio: Diagnostic_studies / Prognostic_studies Límite: Humans Idioma: En Revista: Thromb Haemost Año: 2023 Tipo del documento: Article País de afiliación: Países Bajos

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Hemostáticos / Hemofilia A Tipo de estudio: Diagnostic_studies / Prognostic_studies Límite: Humans Idioma: En Revista: Thromb Haemost Año: 2023 Tipo del documento: Article País de afiliación: Países Bajos
...