Your browser doesn't support javascript.
loading
Intraoperative control by Schanz-screws is inaccurate to achieve the exact amount of correction in de-rotational osteotomies.
Ferner, Felix; Klinder, Annett; Woerner, Michael; Morris, Patrick; Harrer, Joerg; Dickschas, Joerg; Lutter, Christoph.
Afiliación
  • Ferner F; Klinik für Orthopädie, Unfallchirurgie, Handchirurgie, Regiomed Klinik Lichtenfels, Prof. Arneth Str. 2B, 96125, Lichtenfels, Germany. felix.ferner@gmx.de.
  • Klinder A; Orthopädische Klinik und Poliklinik, Universität Rostock, Rostock, Germany. felix.ferner@gmx.de.
  • Woerner M; Osteotomie Komitee, Deutsche Kniegesellschaft, Lichtenfels, Germany. felix.ferner@gmx.de.
  • Morris P; Orthopädische Klinik und Poliklinik, Universität Rostock, Rostock, Germany.
  • Harrer J; Klinik für Orthopädie und Unfallchirurgie, Sozialstiftung Bamberg, Bamberg, Germany.
  • Dickschas J; Klinik für Orthopädie und Unfallchirurgie, Sozialstiftung Bamberg, Bamberg, Germany.
  • Lutter C; Klinik für Orthopädie, Unfallchirurgie, Handchirurgie, Regiomed Klinik Lichtenfels, Prof. Arneth Str. 2B, 96125, Lichtenfels, Germany.
Knee Surg Sports Traumatol Arthrosc ; 31(10): 4319-4326, 2023 Oct.
Article en En | MEDLINE | ID: mdl-37329368
ABSTRACT

PURPOSE:

The accuracy of intraoperative control of correction commonly is achieved by K-wires or Schanz-screws in combination with goniometer in de-rotational osteotomies. The purpose of this study is to investigate the accuracy of intraoperative torsional control in de-rotational femoral and tibial osteotomies. It is hypothesized, that intraoperative control by Schanz-screws and goniometer in de-rotational osteotomies around the knee is a safe and well predictable method to control the surgical torsional correction intraoperatively.

METHODS:

55 consecutive osteotomies around the knee joint were registered, 28 femoral and 27 tibial. The indication for osteotomy was femoral or tibial torsional deformity with the clinical occurrence of patellofemoral maltracking or PFI. Pre- and postoperative torsions were measured according to the method of Waidelich on computed tomography (CT) scan. The scheduled value of torsional correction was defined by the surgeon preoperatively. Intraoperative control of torsional correction was achieved by 5 mm-Schanz-screws and goniometer. The measured values of torsional CT scan were compared to the preoperative defined and intended values and deviation was calculated separately for femoral and tibial osteotomies.

RESULTS:

The surgeon's intraoperative measured mean value of correction in all osteotomies was 15.2° (SD 4.6; range 10-27), whereas the postoperatively measured mean value on CT scan was 15.6 (6.8; 5.0-28.5). Intraoperatively the femoral mean value measured 17.9° (4.9; 10-27) and 12.4° (1.9; 10-15) for the tibia. Postoperatively the mean value for femoral correction was 19.8 (5.5; 9.0-28.5) and 11.3 (5.0; 5.0-26.0) for tibial correction. When considering a deviation of plus or minus 3° to be acceptable femorally 15 osteotomies (53.6%) and tibially 14 osteotomies (51.9%) fell within these limits. Nine femoral cases (32.1.%) were overcorrected, four cases undercorrected (14.3%). Four tibial cases of overcorrection (14.8%) and 9 tibial cases of undercorrection (33.3%) were observed. However, the observed difference between femur and tibia regarding the distribution of cases between the three groups did not reach significance. Moreover, there was no correlation between the extent of correction and the deviation from the intended result.

CONCLUSION:

The use of Schanz-screws and goniometer in de-rotational osteotomies as an intraoperative control of correction is an inaccurate method. Every surgeon performing derotational osteotomies must consider this and include postoperative torsional measurement in his postoperative algorithm until new tools or devices are available to guarantee a better intraoperative accuracy of torsional correction. STUDY

DESIGN:

Observational study. LEVEL OF EVIDENCE III.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Tibia / Fémur Tipo de estudio: Observational_studies Límite: Humans Idioma: En Revista: Knee Surg Sports Traumatol Arthrosc Asunto de la revista: MEDICINA ESPORTIVA / TRAUMATOLOGIA Año: 2023 Tipo del documento: Article País de afiliación: Alemania

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Tibia / Fémur Tipo de estudio: Observational_studies Límite: Humans Idioma: En Revista: Knee Surg Sports Traumatol Arthrosc Asunto de la revista: MEDICINA ESPORTIVA / TRAUMATOLOGIA Año: 2023 Tipo del documento: Article País de afiliación: Alemania
...