Impact of blood collection devices and mode of transportation on peripheral venous blood gas parameters.
Bowen, Raffick A R; Esguerra, Vanessa; Arboleda, Estrella; Cheng, Phil; Hu, Bing.
Afiliación
  • Bowen RAR; Department of Pathology, Stanford Health Care, Stanford, CA 94305, United States. Electronic address: rbowen@stanfordhealthcare.org.
  • Esguerra V; Stanford Hospital and Clinics, Stanford, CA 94305, United States.
  • Arboleda E; Stanford Hospital and Clinics, Stanford, CA 94305, United States.
  • Cheng P; Stanford Hospital and Clinics, Stanford, CA 94305, United States.
  • Hu B; Stanford Hospital and Clinics, Stanford, CA 94305, United States.
Clin Chim Acta ; 548: 117464, 2023 Aug 01.
Article en En | MEDLINE | ID: mdl-37399883
ABSTRACT

BACKGROUND:

Peripheral venous blood (PVB) gas analysis has become an alternative to arterial blood gas (BG) analysis in assessing acid-base balance. This study aimed to compare the effects of blood collection devices and modes of transportation on peripheral venous BG parameters.

METHODS:

PVB-paired specimens were collected from 40 healthy volunteers into blood gas syringes (BGS) and blood collection tubes (BCT), transported by either a pneumatic tube system (PTS) or human courier (HC) to the clinical laboratory, and compared using a two-way ANOVA or Wilcoxon signed-rank test. To determine clinical significance, the PTS and HC-transported BGS and BCT biases were compared to the total allowable error (TEA).

RESULTS:

PVB partial pressure of oxygen (pO2), fractional oxyhemoglobin (FO2Hb), fractional deoxyhemoglobin (FHHb), and oxygen saturation (sO2) showed statistically significant differences between BGS and BCT (p < 0.0001). Compared to HC-transported BGS and BCT, statistically significant increases in pO2, FO2Hb, sO2, oxygen content (only in BCT) (all p < 0.0001), and base excess extracellular (only in BCT; p < 0.0014) concentrations and a statistically significant decrease in FHHb concentration (p < 0.0001) were found in BGS and BCT delivered by PTS. The biases between PTS- and HC-transported BGS and BCT exceeded the TEA for many BG parameters.

CONCLUSIONS:

Collecting PVB in BCT is unsuitable for pO2, sO2, FO2Hb, FHHb, and oxygen content determinations.