New keratoconus staging system based on OCT.
Lu, Nan-Ji; Hafezi, Farhad; Koppen, Carina; Alió Del Barrio, Jorge L; Aslanides, Ioannis M; Awwad, Shady T; Ní Dhubhghaill, Sorcha; Pineda, Roberto; Torres-Netto, Emilio A; Wang, Lin; Chen, Shi-Hao; Cui, Le-Le; Rozema, Jos J.
Afiliación
  • Lu NJ; From the Faculty of Medicine and Health Sciences, University of Antwerp, Wilrijk, Belgium (Lu, Koppen, Ní Dhubhghaill, Rozema); National Clinical Research Center for Ocular Diseases, Eye Hospital, Wenzhou Medical University, Wenzhou, China (Lu, Hafezi, Aslanides, Chen, Cui); State Key Laboratory of Ophthalmology, Optometry and Visual Science, Eye Hospital, Wenzhou Medical University, Wenzhou, China (Lu, Chen, Cui); Department of Ophthalmology, West China Hospital, Sichuan University, Chengdu, Si
J Cataract Refract Surg ; 49(11): 1098-1105, 2023 11 01.
Article en En | MEDLINE | ID: mdl-37531392
ABSTRACT

PURPOSE:

To establish a numerical spectral-domain optical coherence tomography (SD-OCT)-based keratoconus (KC) staging system and compare it with existing KC staging systems.

SETTING:

Eye Hospital of Wenzhou Medical University, Wenzhou, China. DESIGNS Retrospective case-control study.

METHODS:

Scheimpflug tomography, air-puff tonometry, and SD-OCT were performed on 236 normal and 331 KC eyes. All SD-OCT-derived parameters of the corneal epithelium and stroma were evaluated based on their receiver operating characteristic (ROC) curves, area under the curve (AUC), sensitivity, and specificity to discriminate between normal and KC eyes. The best performing parameters were subsequently used to create an OCT-based staging system, which was compared with existing tomographic and biomechanical staging systems.

RESULTS:

236 eyes from 236 normal patients and 331 eyes from 331 KC patients of different stages were included. The highest ranked AUC ROC SD-OCT parameters, derived from stroma and epithelium, were stroma overall minimum thickness (ST AUC 0.836, sensitivity 90%, specificity 67%) and epithelium overall SD (EP AUC 0.835, sensitivity 75%, specificity 78%). A numerical SD-OCT staging system called STEP including 2 parameters-"ST" and "EP"-with 5 stages was proposed.

CONCLUSIONS:

The new SD-OCT-based KC staging system is the first to take the epithelium with its sublayer stroma information into account, showing a strong agreement to the existing staging systems. This system could be incorporated into daily practice, potentially leading to an overall improvement in KC treatment and follow-up management.
Asunto(s)

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Epitelio Corneal / Queratocono Tipo de estudio: Observational_studies / Prognostic_studies Límite: Humans Idioma: En Revista: J Cataract Refract Surg Asunto de la revista: OFTALMOLOGIA Año: 2023 Tipo del documento: Article País de afiliación: Eslovenia

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Epitelio Corneal / Queratocono Tipo de estudio: Observational_studies / Prognostic_studies Límite: Humans Idioma: En Revista: J Cataract Refract Surg Asunto de la revista: OFTALMOLOGIA Año: 2023 Tipo del documento: Article País de afiliación: Eslovenia